14
17.7 Connect the bottom black control wire to T.B.
Terminal #4. Connect the top black wire to T.B.
Terminal #5.
17.8 Do not replace knob until adjusted properly.
Plug unit into electrical outlet. Fill vessel to proper
water level, tighten door and start Autoclave Cycle.
Now proceed with Autoclave Temperature Control
Adjustment per Section 12. Also check for fitting
leaks while unit is under pressure.
18.0 AUTOCLAVE AND DRYER ELEMENT
REPLACEMENT: For the Autoclave Element you
must remove the bottom access cover, shelf screw
and shelf.
18.1 Remove the nut, lockwasher and wire from
each Element terminal.
18.2 Remove the two large nuts and lockwashers
which secure the Element to the vessel.