SFRA45 User Manual
10-2
10.1
(LCR) Screen Display Areas
The main area shows “frequency”, “impedance”, “phase”
and “level” measurements as follows:
“
Frequency”:
is the frequency the generator is currently
running at.
“
Impedance”
is the value (Ω) measured as a ratio of the
voltage and phase angles.
“
Phase”
is the phase angle in degrees between the
voltage and current phases.
“
Level”
is the rms voltage and current levels measured.
The other areas are:
‘
GENERATOR
’ area: Displays the generator frequency, the
output level and if the output is set to ‘on’
‘
INPUT
’ area: Displays the input channel selected. When
the INPUT displays anything but ‘high’ it is in the
‘terminate’ mode. The inputs available are, high Z, 50Ω 18
W, low level and balanced.
‘
MEASURE
’ area: Displays the measurement impedance
reference and the speed of the measurement update.