SFRA45 User Manual
8-12
extend by approximately 50%. This gives the following
ranges:
range reference
nominal full scale
wideband selective
1
3mV
3.16mV
5mV
2
10mV
10mV
15mV
3
30mV
31.6mV
50mV
4
100mV
100mV
150mV
5
300mV
316mV
500mV
6
1V
1V
1.5V
7
3V
3.16V
5V
8
10V
10V
10V
The maximum input signal that can be accurately
measured varies with frequency:
frequency
max input level
MHz
peak V
rms V
1
10
7
5
6
4
10
3
2
15
2
1.4
20
1.5
1
25
1.2
0.8
30
1
0.7
45
0.75
0.5
The input ranges may be selected manually, or by auto
ranging (default). The start range for auto ranging may be
selected if it is known that the signal will not be below a
certain level.