background image

b.

Refill conductivity cell with same standard. 

c.

Press              or               , then 

press               , “

CAL

” icon will 

appear on the display (see Fig. 5). 

d.

Press            or                 to step the displayed value toward the 

standard’s value or hold a key down to cause rapid scrolling of the
reading.

e.

Press             once to confirm new value and end the calibration 

sequence for this particular solution type.

If another solution type is also to be measured, change solution type now
and repeat this procedure.

2.  User Calibration Conductivity/TDS

Instrument must be in 

U S E R

  mode,  see  Section  V.  Solution  Selection,

page 11.

a.

Rinse conductivity cell three times with your standard.

b.

Refill conductivity cell with same standard. 

c.

Press             or               , then press 

 twice in COND/three 

times in TDS. The “

CAL

” icon will appear on the display. 

d.

Press             or               to step the displayed value toward the 

standard’s value or hold a key down to cause rapid scrolling of the
reading.

e.

Press             once to confirm new value and end the calibration 

sequence for this particular solution type.

3.  Resistivity Calibration

Resistivity is the reciprocal of Conductivity.  Resistivity is calibrated only if
conductivity is calibrated for the same solution type. 
16

M S

MR

COND

T D S

CAL

MCLR

CAL

MCLR

COND

T D S

CAL

MCLR

M S

MR

CAL

MCLR

°C

CAL

µ S

COND

4 4 2

Figure 5

4.  Reloading Factory Calibration (Cond or TDS)

If calibration is suspect or known to be wrong, and no standard solution is
available,  the  calibration  value  can  be  replaced  with  the  original  factory
value for that solution. This “FAC”  value is the same for all Ultrameters,
and returns you to a known state without solution in the cell. The “FAC”
internal electronics calibration (which bypasses the electrodes and cell) is
not intended to replace calibration with conductivity standard solutions. If
another  solution  type  requires  resetting,  change  solution  type  and
repeat this procedure.

a.

Press              or               .  

b.

Press            .  (If in USER solution mode, press CAL key twice if 

in Conductivity, and three times if in TDS to skip over tempco and
ratio adjustments.)

c.

Press              key until “

FAC

” appears and release.

d.

Press            to accept the factory calibration setting.  

5.  pH Calibration

Important

:    Always  “zero”  your  Ultrameter  with  a  pH  7  buffer  solution

before adjusting the gain with acid or base buffers, i.e., 4 and/or 10, etc.

a.  pH Zero Calibration

1.

Rinse sensor well 3 times with 7 buffer solution.

2.

Refill sensor well with 7 buffer solution.

3.

Press              to verify the pH 

calibration. If the display 
reads 7.0, skip the pH Zero 
Calibration and proceed to 
section b. pH Gain Calibration.

4.

Press              to enter calibration mode. The “

CAL

”, “

BUFFER

 

and “

7

annunciators will appear (see Figure 6). Displayed value

17

pH

CAL

MCLR

CAL

MCLR

COND

T D S

CAL

MCLR

M S

Figure 6

pH

BUFFER

CAL

Summary of Contents for Ultrameter 6P

Page 1: ...Ultrameter Operation Manual Model 6P 9 00 MYRON L COMPANY...

Page 2: ...ion factor and conductivity TDS ratio Displayed here Temperature readout USER temperature compensation or conductivity TDS ratio Memory Storage Recall pH Calibration ORP Electrode pH Glass Electrode p...

Page 3: ...A reasonable service charge will be made for diagnosis or repairs due to normal wear abuse or tampering This warranty is limited to the repair or replacement of the Ultrameter only The Myron L Company...

Page 4: ...ing Temperature Compensation 13 3 User Programmable Conductivity to TDS Ratio 13 4 VI CALIBRATION 14 A Calibration Intervals 14 B Rules for Calibration in the Ultrameter 14 1 Calibration Steps 15 2 Ca...

Page 5: ...On the left of the Ultrameter s liquid crystal display is shown an indicator of the salt solution characteristic used to model temperature compensation of conductivity and its TDS conversion The indic...

Page 6: ...sample enough times to completely replace the storage solution After use the pH ORP sensor well must be refilled with Myron L Storage Solution and the protective cap reinstalled securely ref Maintena...

Page 7: ...pulling up 10 COND RES TDS 2 Rinse sensor well 3 times with sample to be measured Shake out each sample to remove any residual liquid 3 Refill sensor well with sample 4 Press 5 Take reading 6 IMPORTAN...

Page 8: ...or User 4 Press to accept new solution type E Application of USER Solution Type 1 User Programmable Tempco This feature allows you to change your Ultrameter s temperature compensating factor to anoth...

Page 9: ...e calibration procedures The becomes an ACCEPT key At each point pressing accepts the new calibration value and steps you to the next adjustment or out of CAL mode if there are no more steps To bypass...

Page 10: ...442 Figure 5 4 Reloading Factory Calibration Cond or TDS If calibration is suspect or known to be wrong and no standard solution is available the calibration value can be replaced with the original f...

Page 11: ...sensor well by displaying either Acd or bAS 1 The pH calibration mode is initiated by either completion of the pH Zero Calibration or verifying 7 buffer and pressing the twice while in pH measurement...

Page 12: ...ve memory recall or allow to automatically turn off C Clearing a Record Memory Clear After recalling a certain record location press to clear that memory This space will be the place for the next memo...

Page 13: ...checked every 2 weeks to ensure accuracy Measuring some solutions will require more frequent intervals B Calibration Tracking Records To minimize your calibration effort keep records If adjustments yo...

Page 14: ...cement Order model RPR When ordering be sure to include the model and serial number of your instrument to ensure receiving the proper type Complete installation instructions are provided with each rep...

Page 15: ...extend the sensor s useful life Samples containing chlorine sulfur or ammonia can poison any pH electrode If it is necessary to measure the pH of any such sample thoroughly rinse the sensor well with...

Page 16: ...ef pH ORP Sensor Replacement pg 29 pH readings drift or respond 1 Temporary condition due to memory Clean and rejuvenate sensor ref Cleaning slowly to changes in of solution in pH sensor well for long...

Page 17: ...sensor It is available in quarts and gallons Order SSQ or SSG D Soft Protective Case Padded Cordura Nylon carrying case features a belt clip for hands free mobility Model UCC Registered trade mark of...

Page 18: ...the graph below Users wanting to measure natural water based solutions to 1 would have to alter the internal compensation to the more suitable preloaded 442 values or stay close to 25 C Some who have...

Page 19: ...commonly referenced compounds In the LCD display the solution type being used is shown on the left Besides KCl NaCl and 442 there is the USER choice The benefit of USER is that one may enter the tempe...

Page 20: ...ty of an aqueous solution It is also stated as the Hydrogen Ion activity of a solution pH measures the 34 effective not the total acidity of a solution A 4 solution of acetic acid pH 4 vinegar can be...

Page 21: ...P Oxidation Reduction Potential REDOX 1 ORP as an Indicator ORP is the measurement of the ratio of oxidizing activity to reducing activity in a solution It is the potential of a solution to give up el...

Page 22: ...tential REDOX pg 37 Reduce As in ORP TDS Total Dissolved Solids or the Total Conductive Ions in a solution See Conductivity Conversion to TDS pg 32 Tempco Temperature Compensation See Temperature Comp...

Page 23: ...Ultrameter Operation Manual Addendum Models 4P 6P Software Versions 2 03 2 10 2 51 Later See page 5 to determine the version of software of your Ultrameter UMMA10 01 WEB...

Page 24: ...e resistivity reading is greater than 10 Megohms the Ultrameter performs automatic temperature compensation for high purity water As such the maximum possible value that should be displayed for water...

Page 25: ...e Linked i e or 2 Place the Ultrameter in User mode Refer to SOLUTION SELECTION in Operation Manual for selecting the User Mode 3 Press the arrow key until the menu Linc appears Fig 6 Figure 6 MR COND...

Page 26: ...ink should be canceled See above Canceling User Mode calibration Link 3 Once a link has been established for User mode the link will apply to all measurement modes using User solution selection i e TD...

Reviews: