background image

a.

While in “USER” mode, press             . 

b.

Press              twice (to skip over tempco adjustment), and 

RATIO

” will appear  (see Figure 4).

c.

Adjust with 

     or 

        

keys until new conversion 
ratio is displayed.  

d.

Press

           twice (to skip 

over calibration adjustment) 
to accept new conversion ratio. 
You are now ready to measure samples with the new 
conductivity/TDS ratio.

VI.  

CALIBRATION

A.  Calibration Intervals

Generally,  calibration  is  recommended  about  once  per  month  with
Conductivity  or  TDS  solutions.  Calibration  with  pH  solutions  should  be
checked  twice  a  month.  Calibration  of  ORP  is  not  necessary  (ref.
Calibration Intervals, pg. 23).

B.  Rules for Calibration in the Ultrameter

1.  Calibration Steps

a.  Starting Calibration

Calibration is begun by pressing               while measuring Conductivity, 

TDS  or  pH.  Measuring  continues,  but  the  CAL  icon  is  on,  indicating
calibration is now changeable.

The reading is changed with the                and 

    to match the known 

value. The calibration for each of the 4 solution types may be performed
from either conductivity or TDS mode.

14

CAL

MCLR

M S

MR

In these first five sections, you have learned 
all you need to make accurate measurements.
The following sections contain calibration,
advanced operations and technical information.

M S

MR

CAL

MCLR

CAL

MCLR

T D S

Figure 4

TDS

RATIO

User

Depending  on  what  is  being  calibrated,  there  may  be  1,  2  or  3  steps  to
the calibration procedures. 
 

The  

       becomes an “ACCEPT” key. At each point, pressing  

accepts  the  new  calibration  value  and  steps  you  to  the  next  adjustment
(or out of CAL mode if there are no more steps). 
 

To bypass a calibration step, just press            to accept the present value

as is. 

c.  Leaving Calibration

You  know you are finished when the “

CAL

” icon goes out. Pressing any

measurement  key  abandons  changes  not  yet  accepted  and  exits
calibration mode.  
Leaving pH after the 2nd buffer results in the same gain being entered in
place of the 3rd buffer.

2.  Calibration Limits

There are calibration limits. A nominal “FAC” value is an ideal value stored
by  the  factory.  Attempts  to  calibrate  too  far,  up  or  down,  from  there  will
cause  the  displayed  value  to  be  replaced  with  “FAC”.  If  you  accept  it
(press the “Cal” key), you will have the original default factory calibration
for this measurement. The need to calibrate so far out that “FAC” appears

indicates a procedural problem, wrong standard solution, a very dirty cell

cup or a dying pH/ORP sensor (ref. Troubleshooting Chart, pg. 27).

C.  Calibration Procedures 

1.  Conductivity or TDS Calibration

a.

Rinse conductivity cell three times with proper standard (KCl, 
NaCl, or 442) (ref. Cond./TDS Standard Solutions, pg. 28). For 
user calibration see User Calibration Conductivity/TDS, pg. 16.

15

KCl, NaCl or 442

User

Cond 

Gain only   

Tempco, then Gain

Res 

Done in conductivity

Done in conductivity or TDS

TDS 

Gain only 

Tempco, Ratio, then Gain

pH 

7, acid and/or base 

ORP 

Zero set with pH 7 automatically

CAL

MCLR

CAL

MCLR

CAL

MCLR

Summary of Contents for Ultrameter 6P

Page 1: ...Ultrameter Operation Manual Model 6P 9 00 MYRON L COMPANY...

Page 2: ...ion factor and conductivity TDS ratio Displayed here Temperature readout USER temperature compensation or conductivity TDS ratio Memory Storage Recall pH Calibration ORP Electrode pH Glass Electrode p...

Page 3: ...A reasonable service charge will be made for diagnosis or repairs due to normal wear abuse or tampering This warranty is limited to the repair or replacement of the Ultrameter only The Myron L Company...

Page 4: ...ing Temperature Compensation 13 3 User Programmable Conductivity to TDS Ratio 13 4 VI CALIBRATION 14 A Calibration Intervals 14 B Rules for Calibration in the Ultrameter 14 1 Calibration Steps 15 2 Ca...

Page 5: ...On the left of the Ultrameter s liquid crystal display is shown an indicator of the salt solution characteristic used to model temperature compensation of conductivity and its TDS conversion The indic...

Page 6: ...sample enough times to completely replace the storage solution After use the pH ORP sensor well must be refilled with Myron L Storage Solution and the protective cap reinstalled securely ref Maintena...

Page 7: ...pulling up 10 COND RES TDS 2 Rinse sensor well 3 times with sample to be measured Shake out each sample to remove any residual liquid 3 Refill sensor well with sample 4 Press 5 Take reading 6 IMPORTAN...

Page 8: ...or User 4 Press to accept new solution type E Application of USER Solution Type 1 User Programmable Tempco This feature allows you to change your Ultrameter s temperature compensating factor to anoth...

Page 9: ...e calibration procedures The becomes an ACCEPT key At each point pressing accepts the new calibration value and steps you to the next adjustment or out of CAL mode if there are no more steps To bypass...

Page 10: ...442 Figure 5 4 Reloading Factory Calibration Cond or TDS If calibration is suspect or known to be wrong and no standard solution is available the calibration value can be replaced with the original f...

Page 11: ...sensor well by displaying either Acd or bAS 1 The pH calibration mode is initiated by either completion of the pH Zero Calibration or verifying 7 buffer and pressing the twice while in pH measurement...

Page 12: ...ve memory recall or allow to automatically turn off C Clearing a Record Memory Clear After recalling a certain record location press to clear that memory This space will be the place for the next memo...

Page 13: ...checked every 2 weeks to ensure accuracy Measuring some solutions will require more frequent intervals B Calibration Tracking Records To minimize your calibration effort keep records If adjustments yo...

Page 14: ...cement Order model RPR When ordering be sure to include the model and serial number of your instrument to ensure receiving the proper type Complete installation instructions are provided with each rep...

Page 15: ...extend the sensor s useful life Samples containing chlorine sulfur or ammonia can poison any pH electrode If it is necessary to measure the pH of any such sample thoroughly rinse the sensor well with...

Page 16: ...ef pH ORP Sensor Replacement pg 29 pH readings drift or respond 1 Temporary condition due to memory Clean and rejuvenate sensor ref Cleaning slowly to changes in of solution in pH sensor well for long...

Page 17: ...sensor It is available in quarts and gallons Order SSQ or SSG D Soft Protective Case Padded Cordura Nylon carrying case features a belt clip for hands free mobility Model UCC Registered trade mark of...

Page 18: ...the graph below Users wanting to measure natural water based solutions to 1 would have to alter the internal compensation to the more suitable preloaded 442 values or stay close to 25 C Some who have...

Page 19: ...commonly referenced compounds In the LCD display the solution type being used is shown on the left Besides KCl NaCl and 442 there is the USER choice The benefit of USER is that one may enter the tempe...

Page 20: ...ty of an aqueous solution It is also stated as the Hydrogen Ion activity of a solution pH measures the 34 effective not the total acidity of a solution A 4 solution of acetic acid pH 4 vinegar can be...

Page 21: ...P Oxidation Reduction Potential REDOX 1 ORP as an Indicator ORP is the measurement of the ratio of oxidizing activity to reducing activity in a solution It is the potential of a solution to give up el...

Page 22: ...tential REDOX pg 37 Reduce As in ORP TDS Total Dissolved Solids or the Total Conductive Ions in a solution See Conductivity Conversion to TDS pg 32 Tempco Temperature Compensation See Temperature Comp...

Page 23: ...Ultrameter Operation Manual Addendum Models 4P 6P Software Versions 2 03 2 10 2 51 Later See page 5 to determine the version of software of your Ultrameter UMMA10 01 WEB...

Page 24: ...e resistivity reading is greater than 10 Megohms the Ultrameter performs automatic temperature compensation for high purity water As such the maximum possible value that should be displayed for water...

Page 25: ...e Linked i e or 2 Place the Ultrameter in User mode Refer to SOLUTION SELECTION in Operation Manual for selecting the User Mode 3 Press the arrow key until the menu Linc appears Fig 6 Figure 6 MR COND...

Page 26: ...ink should be canceled See above Canceling User Mode calibration Link 3 Once a link has been established for User mode the link will apply to all measurement modes using User solution selection i e TD...

Reviews: