Myron L Ultrameter 6P Operation Manual Download Page 19

to  indicate  a  certain  setpoint  or  minimum  concentration  or  trend.  The
Ultrameter gives the user the capability to take data in “KCl conductivity
units” to compare to older published data, in terms of NaCl or 442, or as
may be appropriate. The Ultrameter can be used to reconcile data taken
with  other  compensation  assumptions,  especially  with  its  ability  to  allow
custom characteristics through the USER mode.

X V.

CONDUCTIVITY CONVERSION to TOTAL 

DISSOLVED SOLIDS (TDS)

Electrical  conductivity  indicates  solution  concentration  and  ionization  of
the  dissolved  material.  Since  temperature  greatly  affects  ionization,
conductivity measurements are temperature dependent and are normally
corrected  to  read  what  they  would  be  at  25°C  (ref. Temperature
Compensation, pg. 29).

A.  How it’s Done

Once  the  effect  of  temperature  is  removed,  the  compensated
conductivity  is  a  function  of  the  concentration  (TDS).  Temperature
compensation of the conductivity of a solution is performed automatically
by  the  internal  processor,  using  data  derived  from  chemical  tables.  Any
dissolved salt at a known temperature has a known ratio of conductivity to
concentration. Tables  of conversion ratios referenced to 25°C have been
published by chemists for decades.

B.  Solution Characteristics

Real world applications have to measure a wide range of materials and
mixtures  of  electrolyte  solutions.  To  solve  this  problem,  industrial  users
commonly  use  the  characteristics  of  a  standard  material  as  a  model  for
their solution, like the KCl favored by chemists for its stability. 

Users  dealing  with  sea  water,  etc.,  use  NaCl  as  the  model  for  their
concentration  calculations.  Users  dealing  with  freshwater  work  with
mixtures  including  sulfates,  carbonates  and  chlorides,  the  three
predominant  components  (anions)  in  freshwater  that  the  Myron  L
Company  calls  “natural  water”.  These  are  modeled  in  a  mixture  called
“442”  which  the  Myron  L  Company  markets  for  use  as  a  calibration
standard, as it does standard KCl and NaCl solutions.

The  Ultrameter  contains  internal  algorithms  for  these  3  most  commonly
referenced compounds. In the LCD display, the solution type being used
is  shown  on  the  left.  Besides  KCl,  NaCl,  and  442,  there  is  the  “USER”
choice.  The  benefit  of  USER  is  that  one  may  enter  the  temperature
compensation  and  TDS  ratio  by  hand,  greatly  increasing  accuracy  of
readings for a specific solution. That value remains a constant for all 
32

measurements,  and  should  be  reset  for  different  dilutions  or
temperatures.

C.  When does it make a lot of difference?

First, the accuracy of temperature compensation to 25°C determines the
accuracy  of  any  TDS  conversion.  Assume  we  have  industrial  process
water  to  be  pretreated  by  R.O.  Assume  it  is  45°C  and  reads  1500  µS
uncompensated. 

1. 

If NaCl compensation is used, an instrument would report 1035 
µS compensated, which corresponds to 510 ppm NaCl.

2. 

If 442 compensation is used, an instrument would report 1024 
µS compensated, which corresponds to 713 ppm 442.

The difference in values is 40%.

In spite of such large error,  some users will continue to take data in the
NaCl  mode  because  their  previous  data  gathering  and  process
monitoring was done with an older NaCl referenced device. 

Those  who  want  true  TDS  readings  that  will  correspond  to  evaporated
weight  will  select  the  correct  Solution  Type.  If  none  of  the  3  standard
solutions  apply,  the  User  mode  must  be  used.  Temperature
Compensation  (Tempco)  and  TDS  Derivation  below,  details  the  USER
mode.

XVI.

TEMPERATURE COMPENSATION (Tempco) 

and TDS DERIVATION

The  Ultrameter  contains  internal  algorithms  for  characteristics  of  the  3
most commonly referenced compounds. In the display,  the solution type
being used is shown on the left. Besides KCl, NaCl, and 442, there is the
“USER”  choice.  The  benefit  of  USER  mode  is  that  one  may  enter  the
tempco  and  TDS  conversion  values  of  a  unique  solution  from  the
keyboard. 

A.  Conductivity Characteristics

When  making  conductivity  measurements,  the  Solution  Selection
determines the characteristic assumed as the instrument reports what a
measured  conductivity  would  be  if  it  were  at  25°C.  The  characteristic  is
represented by the tempco, expressed in %/°C. If a solution of 100 µS at
25°C  increases  to  122  µS  at  35°C,  then  a  22%  increase  has  happened

over  this  change  of  10°C.  The  solution  is  said  to  have  a  tempco  of  2.2

%/°C.  

33

Summary of Contents for Ultrameter 6P

Page 1: ...Ultrameter Operation Manual Model 6P 9 00 MYRON L COMPANY...

Page 2: ...ion factor and conductivity TDS ratio Displayed here Temperature readout USER temperature compensation or conductivity TDS ratio Memory Storage Recall pH Calibration ORP Electrode pH Glass Electrode p...

Page 3: ...A reasonable service charge will be made for diagnosis or repairs due to normal wear abuse or tampering This warranty is limited to the repair or replacement of the Ultrameter only The Myron L Company...

Page 4: ...ing Temperature Compensation 13 3 User Programmable Conductivity to TDS Ratio 13 4 VI CALIBRATION 14 A Calibration Intervals 14 B Rules for Calibration in the Ultrameter 14 1 Calibration Steps 15 2 Ca...

Page 5: ...On the left of the Ultrameter s liquid crystal display is shown an indicator of the salt solution characteristic used to model temperature compensation of conductivity and its TDS conversion The indic...

Page 6: ...sample enough times to completely replace the storage solution After use the pH ORP sensor well must be refilled with Myron L Storage Solution and the protective cap reinstalled securely ref Maintena...

Page 7: ...pulling up 10 COND RES TDS 2 Rinse sensor well 3 times with sample to be measured Shake out each sample to remove any residual liquid 3 Refill sensor well with sample 4 Press 5 Take reading 6 IMPORTAN...

Page 8: ...or User 4 Press to accept new solution type E Application of USER Solution Type 1 User Programmable Tempco This feature allows you to change your Ultrameter s temperature compensating factor to anoth...

Page 9: ...e calibration procedures The becomes an ACCEPT key At each point pressing accepts the new calibration value and steps you to the next adjustment or out of CAL mode if there are no more steps To bypass...

Page 10: ...442 Figure 5 4 Reloading Factory Calibration Cond or TDS If calibration is suspect or known to be wrong and no standard solution is available the calibration value can be replaced with the original f...

Page 11: ...sensor well by displaying either Acd or bAS 1 The pH calibration mode is initiated by either completion of the pH Zero Calibration or verifying 7 buffer and pressing the twice while in pH measurement...

Page 12: ...ve memory recall or allow to automatically turn off C Clearing a Record Memory Clear After recalling a certain record location press to clear that memory This space will be the place for the next memo...

Page 13: ...checked every 2 weeks to ensure accuracy Measuring some solutions will require more frequent intervals B Calibration Tracking Records To minimize your calibration effort keep records If adjustments yo...

Page 14: ...cement Order model RPR When ordering be sure to include the model and serial number of your instrument to ensure receiving the proper type Complete installation instructions are provided with each rep...

Page 15: ...extend the sensor s useful life Samples containing chlorine sulfur or ammonia can poison any pH electrode If it is necessary to measure the pH of any such sample thoroughly rinse the sensor well with...

Page 16: ...ef pH ORP Sensor Replacement pg 29 pH readings drift or respond 1 Temporary condition due to memory Clean and rejuvenate sensor ref Cleaning slowly to changes in of solution in pH sensor well for long...

Page 17: ...sensor It is available in quarts and gallons Order SSQ or SSG D Soft Protective Case Padded Cordura Nylon carrying case features a belt clip for hands free mobility Model UCC Registered trade mark of...

Page 18: ...the graph below Users wanting to measure natural water based solutions to 1 would have to alter the internal compensation to the more suitable preloaded 442 values or stay close to 25 C Some who have...

Page 19: ...commonly referenced compounds In the LCD display the solution type being used is shown on the left Besides KCl NaCl and 442 there is the USER choice The benefit of USER is that one may enter the tempe...

Page 20: ...ty of an aqueous solution It is also stated as the Hydrogen Ion activity of a solution pH measures the 34 effective not the total acidity of a solution A 4 solution of acetic acid pH 4 vinegar can be...

Page 21: ...P Oxidation Reduction Potential REDOX 1 ORP as an Indicator ORP is the measurement of the ratio of oxidizing activity to reducing activity in a solution It is the potential of a solution to give up el...

Page 22: ...tential REDOX pg 37 Reduce As in ORP TDS Total Dissolved Solids or the Total Conductive Ions in a solution See Conductivity Conversion to TDS pg 32 Tempco Temperature Compensation See Temperature Comp...

Page 23: ...Ultrameter Operation Manual Addendum Models 4P 6P Software Versions 2 03 2 10 2 51 Later See page 5 to determine the version of software of your Ultrameter UMMA10 01 WEB...

Page 24: ...e resistivity reading is greater than 10 Megohms the Ultrameter performs automatic temperature compensation for high purity water As such the maximum possible value that should be displayed for water...

Page 25: ...e Linked i e or 2 Place the Ultrameter in User mode Refer to SOLUTION SELECTION in Operation Manual for selecting the User Mode 3 Press the arrow key until the menu Linc appears Fig 6 Figure 6 MR COND...

Page 26: ...ink should be canceled See above Canceling User Mode calibration Link 3 Once a link has been established for User mode the link will apply to all measurement modes using User solution selection i e TD...

Reviews: