background image

C.  An Example of 2 different solution selections and the 

resulting compensation:

How  much  error  results  from  treating  natural  water  as  if  it  were  KCl  at
15°C?

A  tap  water  solution  should  be  compensated  as  442  with  a  tempco  of
1.68 %/°C, where the KCl value used would be 1.90 %/°C. 

Suppose a measurement at 15°C (or 59°F) is 900 microsiemens of true
uncompensated conductivity.  

Using  a  442  correction  of  10  (degrees  below  25)  x  1.68%  indicates  the
solution  is  reading  16.8%  low.  For  correction,  dividing  by  (.832)  yields
1082 microsiemens as a compensated reading. 

A  KCl correction of 10 (degrees below 25) x 1.9% indicates the solution
is  reading  19%  low.  Dividing  by  (.81)  yields  1111  microsiemens  for  a
compensated reading. The difference is 29 out of 1082 = 2.7%.

Chart 1

1.500%

1.600%

1.700%

1.800%

1.900%

2.000%

2.100%

2.200%

2.300%

2.400%

2.500%

Temperature

0      5    10    15    20   25   30    35   40   45    50    55   60

KCl % / °C

% / °C

30

D.  A Chart of Comparative Error:

In the range of 1000 µS, the error using KCl on a solution that should be
compensated as NaCl or as 442, is shown in the graph below.

Users  wanting  to  measure  natural  water  based  solutions  to  1%  would
have  to  alter  the  internal  compensation  to  the  more  suitable  preloaded
“442” values or stay close to 25°C. Some who have standardized to KCl
based  compensation  may  want  to  stick  with  it,  regardless  of  increasing
error  as  you  get  further  from  25°C.  The  Ultrameter  will  provide  the
repeatability  and  convertibility  of  data  needed  for  relative  values  for
process control.

E.  Other Solutions

A  salt solution like sea water or liquid fertilizer acts like NaCl. An internal
correction  for  NaCl  can  be  selected  for  greatest  accuracy  with  such
solutions. Many solutions are not at all similar to KCl, NaCl or 442. A sugar
solution, or a silicate, or a calcium salt at a high or low temperature may
require  a  “User”  value  peculiar  to  the  application  to  provide  readings
close to the true compensated conductivity.  

Clearly,  the  solution  characteristics  should  be  chosen  to  truly  represent

the  actual  water  under  test  for  rated  accuracy  of  ±1%.  Many  industrial

applications have always been relative measurements seeking a number 

31

Chart 2

(2)%

(1)%

0%

1%

2%

3%

4%

5%

6%

7%

0

5

10

15

20

25

30

35

40

45

50

55

Temperature

NaCl error with KCl tempco

442 error with KCl tempco

Summary of Contents for Ultrameter 6P

Page 1: ...Ultrameter Operation Manual Model 6P 9 00 MYRON L COMPANY...

Page 2: ...ion factor and conductivity TDS ratio Displayed here Temperature readout USER temperature compensation or conductivity TDS ratio Memory Storage Recall pH Calibration ORP Electrode pH Glass Electrode p...

Page 3: ...A reasonable service charge will be made for diagnosis or repairs due to normal wear abuse or tampering This warranty is limited to the repair or replacement of the Ultrameter only The Myron L Company...

Page 4: ...ing Temperature Compensation 13 3 User Programmable Conductivity to TDS Ratio 13 4 VI CALIBRATION 14 A Calibration Intervals 14 B Rules for Calibration in the Ultrameter 14 1 Calibration Steps 15 2 Ca...

Page 5: ...On the left of the Ultrameter s liquid crystal display is shown an indicator of the salt solution characteristic used to model temperature compensation of conductivity and its TDS conversion The indic...

Page 6: ...sample enough times to completely replace the storage solution After use the pH ORP sensor well must be refilled with Myron L Storage Solution and the protective cap reinstalled securely ref Maintena...

Page 7: ...pulling up 10 COND RES TDS 2 Rinse sensor well 3 times with sample to be measured Shake out each sample to remove any residual liquid 3 Refill sensor well with sample 4 Press 5 Take reading 6 IMPORTAN...

Page 8: ...or User 4 Press to accept new solution type E Application of USER Solution Type 1 User Programmable Tempco This feature allows you to change your Ultrameter s temperature compensating factor to anoth...

Page 9: ...e calibration procedures The becomes an ACCEPT key At each point pressing accepts the new calibration value and steps you to the next adjustment or out of CAL mode if there are no more steps To bypass...

Page 10: ...442 Figure 5 4 Reloading Factory Calibration Cond or TDS If calibration is suspect or known to be wrong and no standard solution is available the calibration value can be replaced with the original f...

Page 11: ...sensor well by displaying either Acd or bAS 1 The pH calibration mode is initiated by either completion of the pH Zero Calibration or verifying 7 buffer and pressing the twice while in pH measurement...

Page 12: ...ve memory recall or allow to automatically turn off C Clearing a Record Memory Clear After recalling a certain record location press to clear that memory This space will be the place for the next memo...

Page 13: ...checked every 2 weeks to ensure accuracy Measuring some solutions will require more frequent intervals B Calibration Tracking Records To minimize your calibration effort keep records If adjustments yo...

Page 14: ...cement Order model RPR When ordering be sure to include the model and serial number of your instrument to ensure receiving the proper type Complete installation instructions are provided with each rep...

Page 15: ...extend the sensor s useful life Samples containing chlorine sulfur or ammonia can poison any pH electrode If it is necessary to measure the pH of any such sample thoroughly rinse the sensor well with...

Page 16: ...ef pH ORP Sensor Replacement pg 29 pH readings drift or respond 1 Temporary condition due to memory Clean and rejuvenate sensor ref Cleaning slowly to changes in of solution in pH sensor well for long...

Page 17: ...sensor It is available in quarts and gallons Order SSQ or SSG D Soft Protective Case Padded Cordura Nylon carrying case features a belt clip for hands free mobility Model UCC Registered trade mark of...

Page 18: ...the graph below Users wanting to measure natural water based solutions to 1 would have to alter the internal compensation to the more suitable preloaded 442 values or stay close to 25 C Some who have...

Page 19: ...commonly referenced compounds In the LCD display the solution type being used is shown on the left Besides KCl NaCl and 442 there is the USER choice The benefit of USER is that one may enter the tempe...

Page 20: ...ty of an aqueous solution It is also stated as the Hydrogen Ion activity of a solution pH measures the 34 effective not the total acidity of a solution A 4 solution of acetic acid pH 4 vinegar can be...

Page 21: ...P Oxidation Reduction Potential REDOX 1 ORP as an Indicator ORP is the measurement of the ratio of oxidizing activity to reducing activity in a solution It is the potential of a solution to give up el...

Page 22: ...tential REDOX pg 37 Reduce As in ORP TDS Total Dissolved Solids or the Total Conductive Ions in a solution See Conductivity Conversion to TDS pg 32 Tempco Temperature Compensation See Temperature Comp...

Page 23: ...Ultrameter Operation Manual Addendum Models 4P 6P Software Versions 2 03 2 10 2 51 Later See page 5 to determine the version of software of your Ultrameter UMMA10 01 WEB...

Page 24: ...e resistivity reading is greater than 10 Megohms the Ultrameter performs automatic temperature compensation for high purity water As such the maximum possible value that should be displayed for water...

Page 25: ...e Linked i e or 2 Place the Ultrameter in User mode Refer to SOLUTION SELECTION in Operation Manual for selecting the User Mode 3 Press the arrow key until the menu Linc appears Fig 6 Figure 6 MR COND...

Page 26: ...ink should be canceled See above Canceling User Mode calibration Link 3 Once a link has been established for User mode the link will apply to all measurement modes using User solution selection i e TD...

Reviews: