Myron L Ultrameter 6P Operation Manual Download Page 3

FEATURES and SPECIFICATIONS

A.  Features:

•   Superior resolution 4 digit LCD displays full 9999 µS/ppm.
•   Accuracy of ±1% of reading (not merely full scale).
•   All electrodes are internal for maximum protection.
•   Latest 4 electrode cell technology.
•   Waterproof to 3 feet/1 meter.
•   Autoranging conductivity/TDS/resistivity.
•   Prompts for easy pH calibration.
•   Memory saves 20 readings.
•   Factory calibrations stored in microprocessor.
•   3 conductivity/TDS solution conversions preprogrammed into 

microprocessor.

•   USER feature allows:

Programming your own cond/TDS conversion factor.
Programming your own temperature compensation factor.
Disabling temperature compensation.

B.  General Specifications

Display

    

4 Digit LCD

Dimensions (LxWxH)

    

7.7x2.7x2.5 in. 
196x68x64 mm

Weight

    

13.5oz./383g

Case Material

    

VALOX*

Cond/Res/TDS Cell Material

    

VALOX* 

Cond/Res/TDS Cell Capacity

    

0.2 oz./5 ml

pH/ORP Sensor Well Capacity

    

0.04 oz./1.2 ml

Power

    

9V Alkaline Battery

Battery Life

    

>100 Hours/5000 Readings

Operating/Storage Temperature   

32-132°F/0-55°C

Protection Ratings

    

IP67/NEMA 6 (waterproof to 
3 feet/1 meter)

* ™ GE Corp.

Additional information available on our web site at:

www.myronl.com

2

C.  Specification Chart

 

 

 

If,  in  the  opinion  of  the  factory,  failure  was  due  to  materials  or
workmanship,  repair  or  replacement  will  be  made  without  charge.  A
reasonable  service  charge  will  be  made  for  diagnosis  or  repairs  due  to
normal wear, abuse or tampering. This warranty is limited to the repair or

replacement of the Ultrameter only.  The Myron  L Company assumes no

other responsibility or liability.

E.

Ultrameter Models

Resistivity

0-9999 ppm

10-200 ppt

in 5 autoranges

0-9999 µS

10-200 mS

in 5 autoranges

10K

 

- 30M

 ohms

±1% of reading*

±1% of reading*

±1% of reading**

0.01 (<100 ppm)
0.1 (<1000 ppm)
1.0 (>1000 ppm)

0.01 (<100 µS)
0.1 (<1000 µS)
1.0 (>1000 µS)

0.01 (<100K

 ohm  

)

0.1 (<1000K

 ohm  

)

1.0 (>1MΩ

ohm 

)

0.1 °C/F

±.01 pH

±1 mV

±0.1 °C

0-71°C

32 - 160 °F

0 - 9.99%/°C

Auto 

Temperature 

Compensation

Adjustable 

Temperature 

Compensation

Cond/TDS 

Ratios 

Preprogrammed

Resolution

Adjustable 

Cond/TDS Ratio 

Factor

0.20 - 7.99

KCl, NaCl, 442™

Ranges

0-14 pH

±999 mV

Accuracy

±.01 pH

±1 mV

0-71°C

32 - 160 °F

0 - 9.99%/°C

0 - 9.99%/°C

0-71°C

32 - 160 °F

0-71°C

32 - 160 °F

0-71°C

32 - 160 °F

Conductivity

pH

ORP

TDS

Temperature

6P

3

ULTRAMETER MODELS

3P

4P

    6P

PARAMETERS

       pH/ORP/Temp.      Conductivity/TDS     Conductivity/TDS/pH

    

     Resistivity/Temp.     Resistivity/ORP/Temp.

MYRON L COMPANY

            

6115 Corte Del Cedro
Carlsbad, CA 92009
USA
760-438-2021

* up to 100mS/ppt; 100 - 200mS/ppt: ± 1 - 2% typ. 

**10K

 ohms

 - 10M

 ohms

D.  Warranty/Service 

All Myron L Ultrameters have a one-year warranty except for pH sensors
which have a 6-month limited warranty.  If an instrument fails to operate 
properly, see the Troublshooting Chart, pg. 27.  The battery and pH/ORP
sensor are user replaceable.  For other service, return the instrument 
prepaid to the Myron L Company.

Summary of Contents for Ultrameter 6P

Page 1: ...Ultrameter Operation Manual Model 6P 9 00 MYRON L COMPANY...

Page 2: ...ion factor and conductivity TDS ratio Displayed here Temperature readout USER temperature compensation or conductivity TDS ratio Memory Storage Recall pH Calibration ORP Electrode pH Glass Electrode p...

Page 3: ...A reasonable service charge will be made for diagnosis or repairs due to normal wear abuse or tampering This warranty is limited to the repair or replacement of the Ultrameter only The Myron L Company...

Page 4: ...ing Temperature Compensation 13 3 User Programmable Conductivity to TDS Ratio 13 4 VI CALIBRATION 14 A Calibration Intervals 14 B Rules for Calibration in the Ultrameter 14 1 Calibration Steps 15 2 Ca...

Page 5: ...On the left of the Ultrameter s liquid crystal display is shown an indicator of the salt solution characteristic used to model temperature compensation of conductivity and its TDS conversion The indic...

Page 6: ...sample enough times to completely replace the storage solution After use the pH ORP sensor well must be refilled with Myron L Storage Solution and the protective cap reinstalled securely ref Maintena...

Page 7: ...pulling up 10 COND RES TDS 2 Rinse sensor well 3 times with sample to be measured Shake out each sample to remove any residual liquid 3 Refill sensor well with sample 4 Press 5 Take reading 6 IMPORTAN...

Page 8: ...or User 4 Press to accept new solution type E Application of USER Solution Type 1 User Programmable Tempco This feature allows you to change your Ultrameter s temperature compensating factor to anoth...

Page 9: ...e calibration procedures The becomes an ACCEPT key At each point pressing accepts the new calibration value and steps you to the next adjustment or out of CAL mode if there are no more steps To bypass...

Page 10: ...442 Figure 5 4 Reloading Factory Calibration Cond or TDS If calibration is suspect or known to be wrong and no standard solution is available the calibration value can be replaced with the original f...

Page 11: ...sensor well by displaying either Acd or bAS 1 The pH calibration mode is initiated by either completion of the pH Zero Calibration or verifying 7 buffer and pressing the twice while in pH measurement...

Page 12: ...ve memory recall or allow to automatically turn off C Clearing a Record Memory Clear After recalling a certain record location press to clear that memory This space will be the place for the next memo...

Page 13: ...checked every 2 weeks to ensure accuracy Measuring some solutions will require more frequent intervals B Calibration Tracking Records To minimize your calibration effort keep records If adjustments yo...

Page 14: ...cement Order model RPR When ordering be sure to include the model and serial number of your instrument to ensure receiving the proper type Complete installation instructions are provided with each rep...

Page 15: ...extend the sensor s useful life Samples containing chlorine sulfur or ammonia can poison any pH electrode If it is necessary to measure the pH of any such sample thoroughly rinse the sensor well with...

Page 16: ...ef pH ORP Sensor Replacement pg 29 pH readings drift or respond 1 Temporary condition due to memory Clean and rejuvenate sensor ref Cleaning slowly to changes in of solution in pH sensor well for long...

Page 17: ...sensor It is available in quarts and gallons Order SSQ or SSG D Soft Protective Case Padded Cordura Nylon carrying case features a belt clip for hands free mobility Model UCC Registered trade mark of...

Page 18: ...the graph below Users wanting to measure natural water based solutions to 1 would have to alter the internal compensation to the more suitable preloaded 442 values or stay close to 25 C Some who have...

Page 19: ...commonly referenced compounds In the LCD display the solution type being used is shown on the left Besides KCl NaCl and 442 there is the USER choice The benefit of USER is that one may enter the tempe...

Page 20: ...ty of an aqueous solution It is also stated as the Hydrogen Ion activity of a solution pH measures the 34 effective not the total acidity of a solution A 4 solution of acetic acid pH 4 vinegar can be...

Page 21: ...P Oxidation Reduction Potential REDOX 1 ORP as an Indicator ORP is the measurement of the ratio of oxidizing activity to reducing activity in a solution It is the potential of a solution to give up el...

Page 22: ...tential REDOX pg 37 Reduce As in ORP TDS Total Dissolved Solids or the Total Conductive Ions in a solution See Conductivity Conversion to TDS pg 32 Tempco Temperature Compensation See Temperature Comp...

Page 23: ...Ultrameter Operation Manual Addendum Models 4P 6P Software Versions 2 03 2 10 2 51 Later See page 5 to determine the version of software of your Ultrameter UMMA10 01 WEB...

Page 24: ...e resistivity reading is greater than 10 Megohms the Ultrameter performs automatic temperature compensation for high purity water As such the maximum possible value that should be displayed for water...

Page 25: ...e Linked i e or 2 Place the Ultrameter in User mode Refer to SOLUTION SELECTION in Operation Manual for selecting the User Mode 3 Press the arrow key until the menu Linc appears Fig 6 Figure 6 MR COND...

Page 26: ...ink should be canceled See above Canceling User Mode calibration Link 3 Once a link has been established for User mode the link will apply to all measurement modes using User solution selection i e TD...

Reviews: