24
3
4
1
2
N
L
N
L
LINE
L
N
230 V
BROWN
BLUE
YELLOW/GREEN
1
2
3
4
5
6
VIOLET
GREY
BROWN
BLUE
VIOLET
GREY
1
2
3
4
F2
K1
K6
K2
K1
LP
GB
INSTALLATION
4.9. Wall-mounted plate wiring connections
Figure O
Connect the power cable (LINE) to terminal K2, observing the following positions:
L - SUPPLY (BROWN WIRE)
- GROUND (YELLOW/GREEN wire)
N - NEUTRAL (BLUE wire)
• Connect the generator’s power cable to the respective connectors, observing the following positions:
K6 - brown wire and yellow/green wire.
K5
- white wire – black wire – red wire – green wire – purple wire.
Eyelet connectors - Both found near the card (see fi gure).
• Connect the 2 generator communication wires to connector K4, according to the following positions:
C11 - (BLACK wire)
C12 - (RED wire)
NOTE: Place the excess cable under the card..
• Connect the 2 control wires (0.5 mm cross-section) for the “generator on” indicator light (LP) to connector K3 according to the following positions:
S11 (card connector K3) to S1 (indicator light).
S12 (card connector K3) to S2 (indicator light).
•
Figure O
Summary of Contents for RX DC
Page 1: ...RX DC 97070012 rev 001 0 7 2 0 0 7 I Gb ITALIANO 3 ENGLISH 15 ...
Page 2: ......
Page 14: ......