20
e
c
f
i
m
g
g
GB
INSTALLATION
4.4. Installing the double pantograph arm
WARNING!
The arms are supplied secured together by a belt.
This belt should not be removed until the two free ends of the arms have been connected to their corresponding attachments: the extension
arm (already secured to the wall) and x-ray head.
If the belt is loosened before fi xing the arms in place, releasing them abruptly could damage them and the operator risks being injured.
The dimensions of the double pantograph arm are not designed for use without the above-mentioned fastenings.
Figure E
• Take the washer (f) from the kit and position it at point (i) corresponding to the extension arm (c).
• Pass the cable (m) of the pantograph arm (e) through the extension arm (c) so that it comes out of the hole below.
• Install the pantograph arm (e) on the extension arm (c)
NOTE:
The extension arm’s bushing is self-lubricating.
Do not lubricate the pin of the double-pantograph arm.
• Pass the cable (m) through the inside of the extension arm (c) so that it comes out of the wall-mounted plate.
• Install the plugs (g) in the corresponding holes in the extension arm (c).
Figure E
Summary of Contents for RX DC
Page 1: ...RX DC 97070012 rev 001 0 7 2 0 0 7 I Gb ITALIANO 3 ENGLISH 15 ...