INSTALLATION
INSTALLATION
1. General safety precautions
Cefl a Sc - Cefl a Dental Group guarantees the safety, reliability and performance of the equipment under the following
conditions:
•
Installation and technical service is performed by authorized technicians using Cefl a Sc - Cefl a Dental Group original
spare parts.
•
The electrical system in the dental surgery corresponds to standards I.E.C. 60364-7-710;V2 (Standards regarding
electrical systems in a medical environment) or equivalent standards currently in force in the country where the equi-
pment is installed.
•
The place where the x-ray unit is installed must comply with offi cial directives regulating radiation in the country where
the equipment is used.
•
The equipment is operated as directed in the instructions manual..
This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) This device
may not cause harmful interference, and (2) this device must accept any interference received, including interference
that may cause undesired operation.
This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of
the FCC rules. These limits are designed to provide reasonable protection against harmful interference in a residential
installation. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in
accordance with the instructions, may cause harmful interference to radio communications
Changes or modifi cations not expressly approved by the party responsible for compliance could void the user’s autho-
rity to operate the equipment.
2. Packaging
2.1. Dimensions and contents
• DIMENSIONS: 103X53X35 CM
• WEIGHT: 35 KG
• CONTENTS:
- Documentation and guarantee
- Disposable jig for installation
- Kit
- Wall back-plate (if requested)
- Wall-mounted plate
- Generator
- Tube
- Extension
- Collimator
- Double pantograph arm
2.2. Handling and storage
Indications regarding storage, handling and unpacking are given on the outside of the
cardboard packaging.
These indications must be strictly observed.
1
) The package must be kept upright in the direction indicated by the arrows at all times
during handling and storage.
2
) Avoid banging the package.
3
) Keep the package free from damp.
4
) Do not use hooks to handle the package.
5
) A nameplate indicates the required ambient conditions for storage.
a
) temperature from -15° to 50° C.
b
) relative humidity from 10 to 90%
c
) atmospheric pressure from 500 to 1060 hPa.