INSTALLATION
4. Installation
The x-ray unit must be installed by a qualifi ed technician in compliance with the installation instructions given below as regards both the mechanical
and electrical parts.
WARNING!
Always check that the voltage indicated on the generator’s nameplate corresponds to that for the electrical system.
4.1. Positioning the x-ray unit’s structure
Figure POSITIONING
Block the middle clamp in the opening provided (d) according to which of the three positions available the structure needs to be placed in:
A – PRESET CONFIGURATION WITH WALL-MOUNTED STRUCTURE SET HORIZONTALLY TO THE RIGHT.
B - WALL-MOUNTED STRUCTURE SET VERTICALLY DOWNWARD.
C - WALL-MOUNTED STRUCTURE SET HORIZONTALLY TO THE LEFT.
•
Secure the pin to the plate using the srews (f) and washers (e) provided.
•
Figure A
Figure POSITIONING
4.2. Wall-mounted plate for supporting the x-ray unit
Figure A
Determine the position of the x-ray unit by using the INSTALLATION JIG.
• Position the wall-mount plate with the bottom holes at a height of approx. 120 cm from the fl oor.
• First check that the plate is at the correct height and perfectly horizontal and then mark out the four points where it is to be fi xed on the wall. Drill
the holes and install the most suitable type of fastening system according to the characteristics of the wall (see paragraph 3.1).
• Pass the supply cable through one of the holes in the wall-mount plate and fi x this to the wall by partially tightening the top and bottom screws.
•