MTS 793 Controller Hardware
FlexTest IIm Controller Connections
184
Jumper
configurations
The 498 transition module can be configured for a Temposonics
sensor, a 5 V DC encoder, a 12 V DC encoder, a 24 V DC encoder. The
following is a list of the configuration drawings:
•
049-035-301Absolute Encoder
•
049-035-401+5 Volt Encoder
•
049-035-501+12 Volt Encoder
•
049-035-601+24 Volt Encoder
•
049-035-701Temposonics II
•
053-070-201Temposonics III
ADDA mezzanine
cards
The ADDA module requires special mezzanine cards to support
encoders and Temposonics sensors.
mezzanine cards for encoders and Temposonics sensors support four
input channels.
mezzanine cards for multiturn encoders (Heidenhain 417/425) and
Temposonics III position sensors support two channels.)
•
The encoder mezzanine card counts pulses from the encoder to
determine the displacement of an actuator.
•
The Temposonics mezzanine card interprets the pulses from the
Temposonics sensor to determine the displacement of an actuator.
Optional ADDA II
Configuration
For the optional ADDA II configuration (using the Model 493.50 ADDA
II module), each encoder requires installation of a Model 493.59
Universal Encoder mezzanine card on the Model 493.50 ADDA II
module.
The Model 493.80 Encoder Transition module provides up to four
channels of encoder interface between the Model 493.50 ADDA II
module and various incremental, Temposonics, or other digital and
serial data transducers.