MTS 793 Controller Hardware
Hydraulic Configurations
221
Appendix A
Hydraulic Configurations
This section describes how to connect the Model 493.10 Chassis
(FlexTest GT Controllers) and the Model 493.02 Chassis (FlexTest SE)
to a variety of MTS hydraulic configurations.
Use the following connectors and cables to connect the Model 493.10/
02 Chassis to your HPU with a Model 493.07 Converter Box:
Note
See Appendix B: Model 493.07 Converter Box on
page 227
for
jumpering information.
493 Chassis to 493.07
Converter Box
•
15 contact type D female EMI connector at
J25
of the Model 493.73
HPU Transition module in the rear panel of the Model 493.10
Chassis. Backshell–EMI metallized plastic.
•
14 contact type CPC male connector at J1 of the Model 493.07
Converter Box chassis.
•
Cable–24 AWG 10 conductor with overall foil shield, (Carol C0745
or equivalent) with drain wire connected to metallized plastic
backshell at the 493.10 chassis, and pin 14 at the Model 493.07
Controller Box chassis.
493.07 Converter Box to
HPU
•
24 contact type CPC female connector at J25 of the Model 493.07
Converter Box chassis.
•
14 contact type MS connector at J1 of the HPU.
•
Cable–18 AWG 14 conductor with overall foil shield, (Alpha 2248C
or equivalent) with drain wire connected to pin 4 of connector J25
and pin A of the pump connector.
Hydraulic pump
configurations
The following diagrams show how to connect the chassis to a
hydraulic power unit.
•
Single 493 chassis with a 24 V PLC (programmable logic
controller) pump or 505 pump
•
Single 493 chassis with a non-PLC pump
•
Multiple 493 chassis with a non-PLC pump
•
Compatible controllers controlling a pump