Check the Precharge Pressure
1. Connect the charging kit chuck valve to the accumulator valve stem.
Item Description
1
Locknut
2
Valve Stem
3
Close
4
Open
5
Poppet Type Accumulator Valve. To open and close, use locknut.
6
Or
7
Valve Core
8
Core type accumulator valve. To open and close use chuck valve.
9
THESE ITEMS SUPPLIED BY USER
10
Nitrogen Bottle Valve