38
Temposonics
®
TH CANbus ATEX / IECEx / CEC / NEC / KCs / EAC Ex certified / Japanese approval
Operation Manual
COB-ID
Request /
Respond
DLC
Byte
Description
0
1
2
3
4
5
6
7
0x67F
Rx
8
0x22
0x10
0x10
0x01
0x73
0x61
0x76
0x65
Note:
This takes at least 20 ms of time!
0x5FF
Tx
8
0x60
0x10
0x10
0x01
0x00
0x00
0x00
0x00
COB-ID
Request /
Respond
DLC
Byte
0
1
2
3
4
5
6
7
0x67F
Rx
8
0x22
0x11
0x10
0x01
0x6C
0x6F
0x61
0x64
0x5FF
Tx
8
0x60
0x11
0x10
0x01
0x00
0x00
0x00
0x00
5.7.6 SDO store parameter index 1010
Using the store parameter command, all current settings are transferred into permanent memory.
Table 13: Store parameter and sensor response
Table 14: Restore parameters
Table 15: Device properties
Index
Sub-index
Description
Type
Attribute
Default Value
Comment
1005
COB-ID sync
Unsigned 32
rw
0x080
1008
Device name
String
ro
C304
1009
Hardware version release
String
ro
1.00
100A
Software version release
String
ro
1.01
100B
Node-ID
Unsigned 32
ro
0x7F
100C
Guard time
Unsigned 16
rw
0
100D
Life time factor
Unsigned 8
rw
0
100E
COB-ID Guarding Protocol
Unsigned 32
rw
0x700 + Node-ID
100F
Number of SDOs supported
Unsigned 32
ro
0x01
1014
COB-ID EMCY
Unsigned 32
rw
0x080 + Node-ID
1017
Producer heartbeat
Unsigned 16
rw
0
1018
0
Identity object
Unsigned 8
ro
4
1
Vendor-ID
Unsigned 32
ro
0x00000040
MTS Sensor
2
Product code
Unsigned 32
ro
0x43333034
C304
3
Revision number
Unsigned 32
ro
0x00010005
4
Serial number
Unsigned 32
ro
17143876
5.7.7 Restore default parameters index 1011
Using the restore parameter command, all current settings are restored to default values.
5.7.8 Sensor communication default parameter
These parameters are related to the C304 order code configuration type.