User Manual | MTI 401 & 400
13
Cleaning and Disinfection
Upholstery
Upholstery should be cleaned weekly with a mild
liquid soap and water mixture. afterwards, wipe
upholstery with a clean damp cloth.
MTI’s standard policy on disinfecting MTI medical
equipment is to use a 10% sodium hypochlorite
(bleach) and 90% water solution. If stronger or other
disinfecting and cleaning solutions are used, the
surface is at risk to be damaged or possibly cause
discoloring, cracking, or other types of damage to the
surface. See current CdC guideline for disinfection
and Sterilization in healthcare Facilities.
MTI understands that many customers prefer to use a
pre-mixed disinfecting solution in a spray or wipe. It is
MTI’s recommendation not to use sprays because over
time the overspray will cause damage to the surfaces
if not immediately wiped up. MTI recommends hand
towels or wipes to avoid the possibility of overspray.
once the approved contact time has been obtained,
remove and dry excess liquid remaining on the surface.
See “MTI Cleaning Solution recommendation letter”
for more details on disinfecting solutions which have
been tested on MTI’s products. It is available for
download at http://www.mti.net/downloads.aspx in
the product Support bulletins folder, found under the
resources tab on our website. In the U.S. & Canada,
call (888) 520-4998, or International, call (801) 875-
4998 to request a copy.
Painted and Plastic Surfaces
all painted metal parts have been electrostatically
painted and the paint is baked on for durability. These
surfaces should be wiped down with a quality grade
cleaner and a soft cloth or a damp cloth with mild
soap once a week.
Preventative Maintenance
periodically inspect the following areas:
1. all fasteners and pins should be in place and
tightened securely.
2. lubricate pins and moving joints as needed with
lithium grease.
3. power cord should be free of cuts or other visible
damage.
4. all mechanical functions should operate properly.
have an authorized service technician inspect your
equipment once a year.