38
MSI-7300 Operator’s Manual
6.2.1
Standard Print Strings
1
Current
Tension
Fixed output length: 16. Leading zeros suppressed except for the
LSD.
TTTTTTT
sp
UU
sp
MMMMM
crlf
TTTTTTT
– tension data with -sign if necessary.
UU
– units
MMMMM
– tension mode which for “1” is either NET or GROSS
2
Net Tension
Fixed output length:16. Leading zeros suppressed except for the
LSD.
TTTTTTT
sp
UU
sp
NET
spspcrlf
3
Gross
Tension
Fixed output length: 16. Leading zeros suppressed except for the
LSD.
TTTTTTT
sp
UU
sp
GROSS
crlf
4
Tare Weight
Fixed output length:16. Leading zeros suppressed except for the
LSD.
TTTTTTT
sp
UU
sp
TARE
crlf
5
Total
Weight
Fixed output length: 16. Leading zeros suppressed except for the
LSD.
TTTTTTTTT
sp
UU
sp
TTL
crlf
6
Number of
Samples
Totaled
Fixed output length: 16. Leading zeros suppressed except for the
LSD.
spspspspspsp
SSSSSSS
sp
T-CNT
spcrlf
7
Current
Weight
Mode
Net, Gross, Peak, etc
sp
MMMMM
crlf
8/9
Carriage
Return/
Line Feed
Used to add a space between print records.
crlf
Table 6-1. Standard Print Strings
In the string number entry screen, you can enter combinations of the standard print
strings. For example, to get a NET GROSS TARE printout with a space between
records, simply enter “2349.”
Using SCCMP application (ScaleCore configuration management program), custom
output strings are possible. See the ScCMP programming guide for details.
Note
ScCMP programming can be found on the CD included with the
product or can be downloaded from the MSI website.
The serial output is configured as 9600 baud, Xon/Xoff handshaking, no hardware
handshaking, 1 stop bit, no parity. Other baud rates are possible by special order only.