Calibration
29
5.2 Initial Calibration
Use this procedure only if the capacity and count-by (
d
) needs to be modified. The
initial steps of this procedure will totally erase user setups as well as any previous
calibration.
1) Turn the 7300 on.
2) Press the
ZERO
switch and
the
F1
switch simultaneously.
3) The display blinks
rESEt
.
4) To reset all calibration
constants and setup
parameters, press
F1
.
5) The 7300 requests a
confirmation by displaying
Sure?
.
To cancel the Reset press
the
POWER
or the
ZERO
key.
6) To complete the reset, press
F1
. The Calibrate menu
appears. You must now
recalibrate the system.
7) Press
F1
to start the initial
calibration procedure. The
display shows
Unit
. You
select the force units you
wish to calibrate in.
8) Press
F1
to select the
calibration unit.
9) Use the
F2
key to scroll
through the available
calibration units.
10) When the desired unit is
shown, press
F1
.
11) Next, set the capacity in the
selected units. Capacity must
be set no higher than the load
cell rated capacity.
12) Press
F1
to enter the capacity
setting screen. A capacity of
10000 is the initial value.
If 10000 is the desired capacity in
the selected Calibration Unit, press
F1 and skip to step 16.
13) To change the capacity, press
F2
.
Resetting Capacity and Countby (d)
UHVHW
blinking
6YUH"
&$/
8QLW
8QLW
blinking
&$3
blinking
blinking
Error correction: if you input a wrong
value, press
ZERO
to step back one
digit and reenter.
To enter a decimal point, push
POWER
while the digit is blinking.
For very large capacity Dyna-Links
Enter the capacity in weight X1000
and push the
POWER
key twice
(X1000 mode is indicated by the
M
annunciator during data entry).
e.g. For a 100,000 lb 7300, enter
100, then press the POWER key
twice so the
M
indicator is on.
ENTER/SELECT
F1
ENTER/SELECT
F1
EXIT/CANCEL
POWER
To cancel
Reset
F1
ENTER/SELECT
F1
ENTER/SELECT
F1
SCROLL
F2
ENTER/SELECT
F1
ENTER/SELECT
F1
SCROLL
F2
kg
lb
M
kN
Ton
Press
&Hold
0
ZERO
F2
F1
kg
lb
M
kN
Ton
kg
lb
M
kN
Ton
kg
lb
M
kN
Ton
kg
lb
M
kN
Ton
kg
lb
M
kN
Ton
pounds
Tons (US Short)
Metric Tons
(kg X 1000)
kiloNewtons
kilograms
kg
lb
M
kN
Ton
kg
lb
M
kN
Ton
kg
lb
M
kN
Ton
kg
lb
M
kN
Ton
kg
lb
M
kN
Ton
Figure 5-3 Initial Calibration
Summary of Contents for Dyna-Link 2
Page 2: ...MSI 7300 Dyna Link 2 Tension Dynamometer Operator s Manual PN 152160 Rev C...
Page 3: ......
Page 58: ...Appendix 53 Notes...
Page 59: ...54 MSI 7300 Operator s Manual Notes...
Page 60: ......