Ultrasonic milkanalyser
Operation manual
48/92
Example:
When it is suitable to us this possibility of the analyser? For example if you have
a device factory calibrat
ed for Cow milk, Sheep Milk and UHT milk, but yоu need
oftenly to measure camel milk. Using the methods, explained in details in
Appendix Methods you may make a new calibration without need to send the
analyser back to the producer for calibration. Using this procedure you may
make calibrations for most oftenly analysed milk and to write down the exact
calibration name, which will be shown on the display and printed on the printer.
After starting this menu the display shows:
Select Calibr
Cal1: Sheep
Exit Yes Next
There are the following possibilities:
With button
Exit
– to leave the menu.
With button
Yes
– to confirm the chosen for editing calibration name.
With button
Next
– to choose the next calibration name for editing.
If a calibration for change or edit of name is chosen, the display shows
(example):
Cal:Sheep
PreDef Exit Edit
There are the following possibilities:
With button
PreDef
– to choose a calibration name from the list of preliminary
given names.
With button
Exit
– to leave the menu.
With button
Edit
– to edit the new calibration name.
If a name from the preliminary given names list is chosen, the display shows:
Cal1: Sheep
UHT
Exit Yes Next