UNI MP100 User’s Guide
22
7.7
Sensor Specifications and Default Configurations
Sensor
Range
(ppm)
Resolution
(ppm)
Span*
(ppm)
Low
(ppm)
High
(ppm)
STEL
(ppm)
TWA
(ppm)
Panel
Ring
Response
Time t
90
(s)
CO
0-500
1
100
35
200
100
35
Red
15
0-1000
1
100
35
200
100
35
15
0-1999
1
100
35
200
100
35
15
H
2
S
0-50
0.1
25
10
20
15
10
Light
Blue
15
0-100
0.1
25
10
20
15
10
15
0-200
0.1
25
10
20
15
10
15
0-1000
1
25
10
20
15
10
30
NH
3
0-100
1
50
25
50
35
25
Orange
150
0-500
1
50
25
50
35
25
150
Cl
2
0-50
0.1
10
2
5
1
0.5 Orange
30
ClO
2
0-1
0.01
0.5**
0.2
0.5
0.3
0.1 Orange
120
H
2
0-1000
1
100
100
400
400
100 Orange
70
0-2000
1
100
100
400
400
100
70
HCN
0-100
1
10
4.7
5
4.7
4.7 Orange
200
NO
0-250
1
25
25
50
25
25
Orange
30
NO
2
0-20
0.1
5
1
10
1
1
Orange
30
PH
3
0-20
0.01
5
1
2
1
0.3 Orange
60
SO
2
0-20
0.1
5
2
10
5
2
Orange
15
ETO
0-100
0.1
10
2
5
2
1
Orange
120
0-200
0.1
10
2
5
2
1
Orange
120
O
3
0-1
0.02
1**
0.1
0.2
0.1
0.1 Orange
60
HF
0-20
0.1
6**
2
6
6
3
Orange
90
HCl
0-15
0.1
10**
2
5
5
1
Orange
90
CH
3
SH
0-10
0.1
5
2
5
2
0.5 Orange
20
Acetaldehyde
0-20
0.1
5
2
5
2
1
Orange
120
THT
0-50
0.1
10
5
10
5
5
Orange
60
*
The default span setting equals the recommended span gas concentration.
**
Calibration of these sensors requires a gas generator or other special precautions. See TA Note 6 for
recommended procedures and gas sources.
Sensor
Range
(%)
Resolution
(%)
Span*
(%)
Low
(%)
High
(%)
STEL
(%)
TWA
(%)
Panel
Ring
Response
Time t
90
(s)
O
2
0 - 25
0.1
0.0
19.5
23.5
-
-
Dark
Blue
15
0 - 30
0.1
0.0
19.5
23.5
-
-
15
*
Oxygen sensors in MP100 use pure nitrogen or other inert gas for both Span and Bump Test.