MESSAGES AND INFORMATION
APPENDIX A
19
The following table describes messages that are displayed on the X12+ during patient hookup and transmission.
Table of Messages
Message
Solution
Battery power is low. Replace existing battery with a fully charged battery.
Battery power is at 25%.
Battery power is at 50%.
Battery power is at 75%.
Battery power is at 100% (fully charged).
CALL
A call signal has been transmitted.
CH:XX
The transmission channel that has been set for this unit.
MENU
Label over the Up/Right arrow key that indicates access to menu options
OFF
Label over the Enter key that indicates powering the unit off.
‘RA’
RA fail. Check if the lead wire is off or the electrode needs to be replaced.
‘RL’
RL fail. Check if the lead wire is off or the electrode needs to be replaced.
‘LA’
LA fail. Check if the lead wire is off or the electrode needs to be replaced.
‘LL’
LL fail. Check if the lead wire is off or the electrode needs to be replaced.
A combination of
‘RA/…/LL’
More than one limb lead fail or all leads fail. Check the lead wires and electrodes.
‘V1’
V1 fail. Check if the lead wire is off or the electrode needs to be replaced.
‘V2’
V2 fail. Check if the lead wire is off or the electrode needs to be replaced.
‘V3’
V3 fail. Check if the lead wire is off or the electrode needs to be replaced.
‘V4’
V4 fail. Check if the lead wire is off or the electrode needs to be replaced.
‘V5’
V5 fail. Check if the lead wire is off or the electrode needs to be replaced.
‘V6’
V6 fail. Check if the lead wire is off or the electrode needs to be replaced.
A combination of
‘V1, V2, V3, V4, V5,
V6’
More than one chest lead fail. Check the lead wires and electrodes.
Summary of Contents for X12 Plus