SECTION 2
11
When the electrode sites have been identified and prepped, remove the clear electrode covering and apply an
electrode to each of the sites. Secure each electrode by exerting slight pressure around the outer edge and inner ring
of the electrode.
To connect the lead wires, begin with connecting the LL lead wire (Red – labeled LL or F) to the LL electrode.
Connect the next lead wire on the cable to the next electrode. Continue connecting the snap connectors to the
electrodes as positioned on the main cable.
NOTE: QRS morphology may be slightly different from a standard 12-lead ECG due to torso located limb
electrode placement.
Using the Keypad
The keypad is located on the front, lower portion of the X12+. Three keys are available for navigating through the
LCD menu screens, for powering the X12+ on/off, and for sending calls during transmission:
Up/Right
,
Down
,
and
Enter
.
Menu Off
Up/Right Down Enter
Call
= Any Key
Main Menu
The X12+ main LCD menu displays the following information.
Ch:XX
= Transmission channel
Battery Symbol = 100%, 75%, 50%, or 25% battery charge
Menu
= Label over the Up/Right key for access to menu options
Off
= label over the Enter key to power the X12+ off
RL, RA, LL, LA, V1, V2, V3, V4, V5
, and/or
V6
= Leads in fail
CALL
= a call signal has been transmitted
To send a
CALL
signal, press any one of the three keypad keys. A
CALL
indicator will appear on the LCD display
to notify the user that a call signal has been transmitted.