APPENDIX A
38
ECG Transmission Errors
(cont’d.)
LCD MESSAGE
RESOLUTION
Error in reply
Connection made with the inappropriate response from receiving station.
Contact Mortara customer service.
No MODEM in Docking
Station
Move electrocardiograph to a docking station with a modem installed.
DPAC Failure
Contact Mortara customer service.
No ECG available for
transmission
Attempting batch transmission with no records available.
Memory full
Will occur when 256 orders are pending in the electrocardiograph.
Battery Errors
LCD MESSAGE
RESOLUTION
PLEASE CONTACT
CUSTOMER SERVICE
Contact Mortara customer service.
Battery Error
Contact Mortara customer service.
Password Errors
LCD MESSAGE
RESOLUTION
Tech Password Compare
Error
Ensure correct password entry in both locations.
Admin Password Compare
Error
Ensure correct password entry in both locations.
Error: Empty Administrator
Password
Administrative password is required; technician password may be empty.
Configuration Errors
LCD MESSAGE
RESOLUTION
DUPLICATE SITE NUMBERS
Ensure appropriate site number has been entered.
General Errors
LCD MESSAGE
RESOLUTION
Not Properly Docked
Apply slight pressure to top of electrocardiograph ensuring a proper seat in the
docking station.
ELI 10 NOT PROPERLY
DOCKED
Apply slight pressure to top of electrocardiograph ensuring a proper seat in the
docking station.
Battery Too Low for Operation
Return electrocardiograph to docking station and allow to charge for a minimum
of 4 hours.
Summary of Contents for ELI 10 Rx
Page 14: ...USER SAFETY INFORMATION xii...
Page 18: ...GENERAL CARE xvi...
Page 42: ...SECTION 3 18...
Page 68: ...APPENDIX B 44...