SECTION 6
31
Reasons
The reasons statements indicate why a particular interpretive statement was printed. Reasons statements print
enclosed in [square brackets] within the interpretive text if the interpretation option is turned on. Turning the
reasons statement function on or off does not affect the measurements performed or the interpretive statements
selected by the analysis program.
For Example:
Anteroseptal Infarct [40+ MS Q WAVE IN V1-V4]
Where “Anteroseptal Infarct” is the interpretive statement and “40+ MS Q WAVE IN V1-V4” is
the reasons statement or explanation as to why the interpretive statement was printed.
Append
A status or statement phrase can be appended to the ECG. Either “UNCONFIRMED REPORT” or “Reviewed by”
can be selected; however, if you wish to have nothing appended to the ECG, “Blank” may be selected.
Delete Rule
Defines the ECG auto-delete status in the patient directory. ECGs that are marked for deletion will be automatically
removed or erased based on its date (a first in/first out philosophy) to make room necessary for the new ECG record.
ECGs are only erased from the directory when they are marked for deletion and if the directory becomes full. More
than one ECG may be removed from the directory in order to make room for the new incoming record. The delete
rule selections are:
Post Plot = ECG is automatically marked for deletion after printing
Post Transmit = ECG is automatically marked for deletion after transmission
Post Plot/Transmit = ECG is automatically marked for deletion after transmission and printing
Storage Sensitivity
Dictates the resolution of all stored ECG records. The sensitivity setting is either normal or high. If the value is set
to high, the stored ECG will have a high resolution. As a result, the record size will be large and will reduce the
storage capacity in the ECG directory.
Auto-Save ECG
Defines whether or not the ELI 10 Rx automatically saves newly acquired ECGs in the directory once they are
acquired and printed. If the auto-save configuration option is set to NO and the record is printed, the ELI 10 Rx will
prompt you to “Save ECG?”
F1
(SAVE)
will store the ECG in the directory.
NOTE: Manually save ECGs by selecting
F5
(More)
from the acquired ECG view and
F5
(Save)
.
Caps Lock
All character entry is translated to uppercase.
Average RR
Enabling this option will display an averaged RR value on the report
.