Operation 2018-03-22
Non-Autoclavable Components: Wipe with Ethanol
• Components maintained this way:
Probe Cord
Tester
Procedure:
■
Disinfection
Wipe the components with a piece of gauze dampened with Ethanol for Disinfection (Ethanol 70 vol% to 80 vol%).
Do not use anything except Ethanol for Disinfection (Ethanol 70 vol% to 80 vol%). Do not use too
much ethanol as it could seep inside and damage the components.
Do not immerse the components in or wipe it with any of the following: functional water (acidic
electrolyzed water, strong alkaline solution, and ozone water), medical agents (glutaral, etc.),
medicinal agents (glutaral, etc.), or any other special types of water or commercial cleaning liquids.
Such liquids may result in plastic degradation, metal corrosion and adhesion of the residual medical
agent to the components.
Never clean the components with chemicals such as formalin cresol (FC) and sodium hypochlorite.
These will damage the plastic parts of the components. If any of these liquids being applied to the
components, wash it off in running water.