Page11
Operation 2018-03-22
Operating the Unit
1.
Turn the unit on.
2.
Hook the contrary electrode in the corner of the patient's
mouth.
Do not use an ultra sonic scaler with the contrary electrode
attached to the patient.
Electrical noise from the scaler could interfere with canal
measurements.
Make sure that the contrary electrode, file holder etc. do
not come into contact with an electric power source such as
an electrical socket. This could result in a severe
electrical shock.
The contrary electrode could cause an adverse reaction if
the patient has an allergy to metals. Ask the patient about
this before using the contrary electrode.
Take care that medicinal solutions such as formalin cresol
(FC) or sodium hypochlorite do not get on the contrary
electrode or the file holder. These could cause an adverse
reaction such as inflammation.
3. Clip the file holder to the metal shaft of the file.
Always clip the file holder to the upper part of file shaft,
near the handle. The metal and plastic part of the file
holder can be damaged if they are attached to the file’s
cutting part or the transition to the cutting part.
1. Press in direction
of arrow with the
thumb.
2. Clip file.
3. Release thumb.
File Holder.
Press
Contrary
Electrode
Corner of Mouth
File or Reamer
Handle
Metal part of
file holder
Cutting part and
transition to cutting
part.
* Do not clip onto this
part!