Operation 2018-03-22
Page18
7.
Maintenance
There are 3 ways to clean and disinfect components depending on the component.
Be sure to follow the procedure below when performing daily maintenance.
Be careful to avoid cross contamination when performing maintenance.
Autoclavable Components
• Components maintained this way:
File Holder
Contrary Electrode
Long File Holder (option)
Take out the file before cleaning the file holder.
Other than the components listed above, seer “Non-Autoclavable Components: Wipe with Ethanol”
on page 21 for how to disinfect components.
Procedure:
■
Cleaning
1. Disconnect the file holder (or long file holder) and contrary
electrode from the probe cord.
2. Clean them off in running water with a soft brush and then
wipe off the water.
If a medical agent being used for the treatment has
adhered to the components, wash it off in running water.
Do not clean the components with an ultrasonic cleaning
device.
After washing is complete, check to see if the file holder or
long file holder, including its inside, is completely dry. If
any water remains inside the component, expel it with an
air gun or another such tool. Failure to do so could result
in the remaining water coming out during use and cause
malfunction or poor sterilization.
If dust or other impurities adhere to the hook of the file
holder or long file holder, they may cause malfunction.
Do not use the high-temperature washer-disinfector.
Cleaning
Disinfection
Packing
Sterilization
Hook