Operation 2018-10-21
Page20
■
Sterilization
Autoclave the file holder, contrary electrode, and long file holder after use for each patient.
Recommended temperature and time:
+134°C (273.2°F), 6 minutes minimum with a sterilization pouch.
Minimum drying time after sterilization: 10 minutes.
or
Recommended temperature and time:
+121°C (249.8°F), 60 minutes minimum with a sterilization pouch.
Minimum drying time after sterilization: 10 minutes.
To prevent the spread of serious, life-threatening infections such as HIV and hepatitis B, the file
holder, long file holder, and contrary electrode must be autoclaved after each patient’s treatment has
been completed.
The file holder, long file holder, and contrary electrode are extremely hot after autoclaving; do not
touch until they cool off.
Do not sterilize the components by any method other than autoclaving.
Autoclaving and drying temperatures must never 135°C (275°F). Excess temperature could
cause the contra angle to malfunction or could cause discoloration.
Take the file out of the file holder or long file holder before autoclaving.
Clean everything throughly before autoclaving. Any chemicals or foreign debris left on components
could cause them to malfunction or could cause discoloration.
Do not leave the file holder, long file holder, and contrary electrode in the autoclave.
For sterilizing files, follow the manufacturer’s recommendations.