6. Reset to default values
Menu level
Parameterization level
Supporting points - number of additional supporting points,
Default:
In addition to the start and end value, 8 extra supporting points can be defined to
linearise non-linear sensor values. Only the activated set point parameters are
displayed.
Display values for supporting points,
Under this parameter the supporting points are defined on a value basis. At the
sensor calibration one will be asked at the end (like at final value/offset, too), if a
calibration shall be triggered.
Analogue values for setpoints,
…
The set points are only displayed with the works calibration (
4-20 mA
) Here, the
desired analogue values can be freely selected. The input of constantly rising
analogue values must be carried out by the customer/user.
6. Reset to default values
To return the unit to a
defined basic state
, a reset can be carried out to the default values.
The following procedure should be used:
• Switch off the power supply
• Press button [
P
]
• Switch on loop current (approx. 3.8 mA) and press
[P]-
button until
is shown in the
display.
With reset, the default values of the program table are loaded and used for subsequent
operation. This puts the unit back to the state in which it was supplied.
Caution! All application-related data are lost.