SECTION 3 

 

OPERATION 

 

STARTING THE CABINET 

The cabinet is switched on by the green rocker switch to the right of the control 
panel. If the safety screen is fitted the fluorescent light will come on and the fan 
will remain off. The fan will start when the safety screen is removed. 
 

UV CYCLE OPERATION 

 

TO START A UV CYCLE THE SAFETY SCREEN MUST BE IN POSITION.

 

 
 

 
 
 
 
 
 
 
 
 
 
 

STARTING A UV STERILISATION CYCLE 

 

Fit the Safety Screen and turn the cabinet on. 
 
Press the UV button on the membrane panel to select the duration of the cycle. 

• 

1 Press = 5 minutes 

• 

2 Presses = 10 minutes 

• 

3 Presses = 15 minutes 

• 

4 Presses = 20 minutes 

• 

5 Presses = 25 minutes 

• 

6 Presses = 30 minutes 

 
The UV pre-set timed cycle will start, the fluorescent light will switch off and the 
UV lamp will switch on. The red UV warning light on the membrane panel will 
flash to indicate the cycle selected – (1 flash = 5 minutes, 2 flashes = 10 
minutes etc.)  
 
When the UV timing cycle is finished, the UV light will switch off, the UV 
warning light will flash slowly and the audio will beep every 4 seconds to inform 
the cycle has finished. 
 
A cycle may be interrupted at any time by pressing and holding the UV button. 
The audio beep will stop and the cabinet will return to normal mode. 
 

Summary of Contents for PCR650

Page 1: ...ISSUE C 26 th Sept 2008 PCR650 OPERATING AND MAINTENANCE MANUAL ...

Page 2: ...2 ...

Page 3: ...LARM 7 SYSTEM MONITOR 7 SECTION 4 8 FILTERS 8 PRE FILTER CHANGING 8 MAIN HEPA FILTER CHANGING 8 SECTION 5 9 MAINTENANCE 9 FUSES 9 LIGHTING 9 UV TUBE 9 AIRFLOW 9 LOW AIRFLOW ALARM 9 CIRCUIT DIAGRAM 10 SECTION 6 11 SERVICING 11 Warning This system must be used in compliance with these instructions and any repairs or maintenance carried out by qualified personnel For parts or service information plea...

Page 4: ...ean air Laminar Flow over the entire worksurface To further reduce the risk of cross contamination between samples the PCR cabinet is supplied with a time controlled UV light to sterilise the working area between samples 6 individual pre set time functions are provided as standard Circulaire Circulaire Circulaire Circulaire PCR PCR PCR PCR Ca Ca Ca Cabinet binet binet binet External Dimensions 650...

Page 5: ...e sited in a draught free position ASSEMBLY The cabinet is supplied fully assembled with the main HEPA filter and pre filter fitted TESTING COMMISSIONING A test certificate will be supplied for conformity to CE marking electrical test and airflow ...

Page 6: ...n of the cycle 1 Press 5 minutes 2 Presses 10 minutes 3 Presses 15 minutes 4 Presses 20 minutes 5 Presses 25 minutes 6 Presses 30 minutes The UV pre set timed cycle will start the fluorescent light will switch off and the UV lamp will switch on The red UV warning light on the membrane panel will flash to indicate the cycle selected 1 flash 5 minutes 2 flashes 10 minutes etc When the UV timing cycl...

Page 7: ...on again When an audio beep sounds release the buttons The counter will be reset and the system return to normal mode LOW AIRFLOW ALARM The airflow is continuously monitored while the cabinet is running An audible alarm will sound if the airflow drops below a preset level and the red low airflow warning light will flash on the control panel This is an indication that the Pre Filter requires replac...

Page 8: ...rm is activated Remove the thumbscrews on top of the cabinet and lift off the pre filter retainer Replacement pre filters may be obtained from Monmouth Scientific Part No PF 0023 MAIN HEPA FILTER CHANGING The HEPA filter should be changed by trained personnel Contact Monmouth Scientific for advice ...

Page 9: ...l panel and may be changed by removing the two securing screws from inside the cabinet and hinging the control panel open The UV light fuse 2A Type F is located on the control circuit board AIRFLOW WARNING This procedure requires adjustment of live circuits and should only be carried out by a qualified person The airflow should be set using a calibrated rotating vane anemometer The fan speed is ad...

Page 10: ...10 CIRCUIT DIAGRAM ...

Page 11: ...s Check replace pre filter Check airflow monitor and re calibrate if necessary Check safety interlocks and alarms Inspect electrical components lighting cables etc Issue test report and airflow certificate For parts or service information please contact Monmouth Scientific on 44 0 1278 458090 ...

Page 12: ...12 Monmouth Scientific Units 5 6 Kilnside East Quay Bridgwater Somerset TA6 4DB Email info monmouthscientific co uk Monmouth Scientific is a trading division of Monmouth Surgical Ltd ...

Reviews: