11
SECTION 6
SERVICING
An annual service is recommended and will include the following points:
•
Check / replace pre-filter
•
Check airflow monitor and re-calibrate if necessary
•
Check safety interlocks and alarms
•
Inspect electrical components, lighting, cables etc.
•
Issue test report and airflow certificate.
For parts or service information please contact Monmouth Scientific on:
+44 (0) 1278 458090