Moving Parts Safety
The instrument contains moving parts that can cause injury. Under normal conditions, the
instrument is designed to protect you from these moving parts.
WARNING!
If the instrument is used in a manner not specified by Molecular
Devices, the protection provided by the equipment might be impaired.
To prevent injury:
Never try to exchange labware, reagents, or tools while the instrument is operating.
Never try to physically restrict the moving components of the instrument.
Keep the instrument work area clear to prevent obstruction of the movement. Provide
clearance in front of the instrument of 18 cm (7.1 in.) for the plate drawer.
CAUTION!
To prevent damage to the instrument, the height of the plate must not
exceed 17 mm, including the lid if the plate is lidded.
WARNING!
Do not attempt to access the interior of the instrument unless
specifically instructed to do so. The moving parts inside the instrument can cause
injury. Do not operate the instrument with any covers or panels removed.
Note:
Observe all warnings and cautions listed for all external devices attached to or in
use during the operation of the instrument. See the applicable user guide for the
operating and safety procedures of that device.