Problems during Measurement
Problem
Cause
Remedy
The system does not build up
pressure during a measurement
- The package is leaking around
the needle and or septum
- The Pressure Rise setting may
be too high
- The package is very porous
- No compressed air supply
- Check the connections
- Switch on printer
- Switch printer ON line
Package is not inflated fully even
though a pressure rise is
displayed
- Clogged needle
- Stop measurement and clean
needle
Pressure loss too high in a
package
- The package is leaking around
the needle
- Poor adhesion of the septum
caused by a dirty package
surface
- Leaks in needle/test head
caused by defective or porous
O-rings
- Check settings of printer (see
manual)
Measurement does not terminate
automatically after bursting of
the package
- Auto-Stop Threshold value set
too high
- Set Auto-Stop Threshold value
to a lower value
Summary of Contents for Dansensor Lippke 4000
Page 1: ...P N 330654 E 01 2018 Package Test System Lippke 4000 4500 User Guide EN...
Page 3: ...P N 330654 E 01 2018 COPYRIGHT Dansensor A S 2 User Guide EN Lippke 4000 4500...
Page 7: ...P N 330654 E 01 2018 COPYRIGHT Dansensor A S 6 User Guide EN Lippke 4000 4500...
Page 11: ...P N 330654 E 01 2018 COPYRIGHT Dansensor A S 10 User Guide EN Lippke 4000 4500...
Page 17: ...P N 330654 E 01 2018 COPYRIGHT Dansensor A S 16 User Guide EN Lippke 4000 4500...
Page 51: ...P N 330654 E 01 2018 COPYRIGHT Dansensor A S 50 User Guide EN Lippke 4000 4500...
Page 75: ...P N 330654 E 01 2018 COPYRIGHT Dansensor A S 74 User Guide EN Lippke 4000 4500...
Page 77: ...P N 330654 E 01 2018 COPYRIGHT Dansensor A S 76 User Guide EN Lippke 4000 4500...