How does it work?
1. Cut out the sample
from the package. The sample should have a size of app. 60 x 60 mm (must
cover the O-ring
) and the valve must be placed in the centre of the sample.
2. Turn the selector knob
to the “VTU TEST” position.
3. Release the handle
and open the transparent lid
.
4. Place the sample in the device as shown making sure that the valve
is centred inside the O-ring
:
- To perform a valve “Burst” test, place the sample with the “outside” surface oriented up.
- To perform a valve “Leak” test, place the sample with the “outside” surface oriented down.
NOTE! The “outside” surface of the sample is the side that was on the exterior surface
of the package before cutting out the sample.
5. Close the lid
and lock the handle
.
6. Perform the test.
When you need to bypass the VTU unit to perform tests on packages using the auxiliary connected
equipment the selector knob
must be turned to the “PACKAGE TEST” position.
3
4
5
1
6
2
Summary of Contents for Dansensor Lippke 4000
Page 1: ...P N 330654 E 01 2018 Package Test System Lippke 4000 4500 User Guide EN...
Page 3: ...P N 330654 E 01 2018 COPYRIGHT Dansensor A S 2 User Guide EN Lippke 4000 4500...
Page 7: ...P N 330654 E 01 2018 COPYRIGHT Dansensor A S 6 User Guide EN Lippke 4000 4500...
Page 11: ...P N 330654 E 01 2018 COPYRIGHT Dansensor A S 10 User Guide EN Lippke 4000 4500...
Page 17: ...P N 330654 E 01 2018 COPYRIGHT Dansensor A S 16 User Guide EN Lippke 4000 4500...
Page 51: ...P N 330654 E 01 2018 COPYRIGHT Dansensor A S 50 User Guide EN Lippke 4000 4500...
Page 75: ...P N 330654 E 01 2018 COPYRIGHT Dansensor A S 74 User Guide EN Lippke 4000 4500...
Page 77: ...P N 330654 E 01 2018 COPYRIGHT Dansensor A S 76 User Guide EN Lippke 4000 4500...