![Mocon AMETEK Dansensor Lippke 5000 User Manual Download Page 115](http://html1.mh-extra.com/html/mocon/ametek-dansensor-lippke-5000/ametek-dansensor-lippke-5000_user-manual_1816852115.webp)
Dansensor
®
Lippke
®
5000
EN
User Guide
113
4. The
System Leak Test
screen shows the date, test pressures and measured pressure loss
for the most recent test.
Press
Perform test
.
5. The test screen shows the parameters for the current test. The test pressure (P) is
automatically adjusted to 1000 mbar below the current inlet pressure and the leak limit
(ΔP) is defined as follows:
Test pressure below 2500 mbar -> 5 mbar
Test pressure 2500 to 5500 mbar -> 0.2% of test pressure
Press the
button to start the test.
6. The test goes through the various stages...
...and when finished the test result is displayed.
The highest measured pressure drop is displayed and for failed tests the specified time
shows how long it took before the specified leak limit (ΔP) was exceeded.
Summary of Contents for AMETEK Dansensor Lippke 5000
Page 1: ...P N 380419 C 05 2022 UserGuide...
Page 2: ...This blank page has been inserted to enable double sided printing of the document...
Page 4: ...2 User Guide EN Dansensor Lippke 5000 P N 380419 C 05 2022...
Page 38: ...36 User Guide EN Dansensor Lippke 5000 P N 380419 C 05 2022...
Page 88: ...86 User Guide EN Dansensor Lippke 5000 P N 380419 C 05 2022...
Page 106: ...104 User Guide EN Dansensor Lippke 5000 P N 380419 C 05 2022...
Page 128: ...126 User Guide EN Dansensor Lippke 5000 P N 380419 C 05 2022...
Page 135: ...This blank page has been inserted to enable double sided printing of the document...
Page 136: ...0 0 VSPQF 4 3 OOFEFWFK 3JOHTUFE FONBSL 5FM JOGP EL NPDPO BNFUFL DPN XXX BNFUFLNPDPO DPN...