STEP 3: Configure ACI Modules
Rev. 1.5 -- Use or Disclosure of Data Contained on this Page is Subject to the Copyright Restrictions on the Inside Title Page of this Document
187
Wired ACIs
2.
If configuring a wired ACI Module, select
Wired
from the
Comm Mode
drop
down menu. The Module number (1-4) will be the Module ID assigned to the
wired ACI device.
3.
Identify any EMG Channels from the
Input Type
drop down list.
Configuring wired ACI modules unlocks EMG specific configuration options.
Wireless IMUs
4.
If configuring a wireless IMU, select
Wireless IMU
from the
Comm Mode
drop down menu and right or left IMU from the
Input Type
drop down menu.
5.
Click the
Refresh Discovered Device List
button to search for wireless
IMUs.
6.
Drag and drop the selected device from the Discovered Devices list into the
appropriate “Set #1” or “Set #2”
Wireless Device
box.
Configuring a New ACI for the First Time
To configure a new ACI perform the following steps:
Input Type (Wired)
This field is used to identify which inputs on an ACI
are EMG sensors.
The possible values are:
•
None
•
CH 1
•
CH 2
•
CH 3
•
CH 4
Refresh Discovered Device
List
Updates the list of discovered devices.
Discovered Devices
Displays list of discovered devices.
Table 23. Arm Control Interface - Fields and Buttons
Fields/Buttons
Description
If configuring the system to have two sets of IMUs, be sure to clearly label each
set of IMUs for the client.
Assigning an ACI Module ID is only required if more than one ACI Module is
being used. Note that all ACI Modules are shipped with a default ID of 1. In
almost every case, the ACI should be assigned to Module 1 in Step 3.