4
Preparing the Stabil-Ion System for Operation
4-36
Series 370 Stabil-Ion Instruction Manual - 370119
FPS
Definition:
Front Panel Status. Displays the status of the current tube setup condition.
Modifiers:
None
Response:
Example:
If the Controller is set as follows: for IG1 – gas a, pressure range low, single filament
operation, filament #1; and for IG2 – gas b, pressure range high, both filaments
operating CGA gas b, CGB gas b, then the response would be:
0
,
0, 0
,
0, 1, 1, 1, 0,1,1 CRLF
SWS
Definition:
Display status of gauge control switches/remotes.
Modifiers:
None
Response:
Example:
If the Controller is set for pressure range 1 low, pressure range 2 high, filament 1 #1,
and filament 2 #2, then the response would be:
0,1,0,1 CRLF
PCS
Definition:
Display process control channel status.
Modifiers:
1
or
2
or
3
or
4
or
5
or
6
or
B
or none.
Response:
Depends on modifier:
0
1
IG1
gas
a
b
pressure range
L
H
filament operation
single
both
filament #
1
2
IG2
gas
a
b
pressure range
L
H
filament operation
single
both
filament #
1
2
CGA
gas
a
b
CGB
gas
a
b
0
1
IG1 filament #
1
2
IG2 filament #
1
2
IG1 pressure range
L
H
IG2 pressure range
L
H
Summary of Contents for Stabil-Ion 370 Series
Page 2: ......
Page 8: ...Table of Contents iv Series 370 Stabil Ion Instruction Manual 370119...
Page 36: ...2 Initial Setup Procedures 2 16 Series 370 Stabil Ion Instruction Manual 370119...
Page 46: ...3 Installation 3 10 Series 370 Stabil Ion Instruction Manual 370119...
Page 58: ...3 Installation 3 22 Series 370 Stabil Ion Instruction Manual 370119...
Page 121: ......