Preparing the Stabil-Ion System for Operation
4-5
Series 370 Stabil-Ion Instruction Manual - 370119
4.3.2 Loading the
Stabil-Ion Memory Module
Before the system is ready to operate, the calibration data in the
Memory Module needs to be loaded into the Controller and the correct
gauge location selected at power up.
1.
Before turning Power ON, decide which Gauge/gas table location
will contain data from the specific Stabil-Ion Gauge Memory
Module. The default is
IG1 GAS a
.
Verify that the Gauge is
connected to that IG channel. Use Table 4-2 on page 4-6 to
specify gauge location and gas type.
2.
With the Controller powered OFF, install the Memory Module
(see Figure 4-4).
Figure 4-3b
Electrometer Function Input Menus Diagram
Figure 4-4
Electrometer Module with
Memory Module Installed
Protect the Memory Module from static electricity. Touch a grounded,
unpainted metal surface on the 370 Control Unit and then touch the surface
of the anti-static bag before removing the Memory Module from its bag.
Information Menu
CANCEL
SELECT
SELECT
IG serial No.'s
IG1 a, b
IG2 a, b
temporary
swap area
swap
be sure!
CANCEL
CANCEL
swap TMP
&
IG1 a
SELECT
swap TMP
& IG1 b
hit select
to swap
SELECT
updating
changes
SELECT
swap TMP
& IG2 a
swap TMP
& IG2 b
hit select
to swap
updating
changes
hit select
to swap
updating
changes
hit select
to swap
updating
changes
CG serial
No.'s
*
CGA (a)
CGB (b)
temporary
swap area
swap
be sure!
swap TMP
&
CGA (a)
SELECT
SELECT
SELECT
swap TMP
& CGB (b)
hit select
to swap
updating
changes
hit select
to swap
updating
changes
Turnoff
History
Newest
Event
0
Power On
Start No.
Oldest
Event No.
Code Version
No.
IG1,2 Off
Cmd
#
Pwr Sup 1,2
Off
#
> Ovp
Thld
#
Setup Menu
(see previous page)
Status for IG1
(see previous page)
SELECT
SELECT
* CG Serial No.'s storage only applies to
calibrated
CONVECTRON
Gauges with
Memory Modules.
Memory Module
IG1 L F1
N2 TORR
▲
▼
Cancel
Select
ELECTROMETER
370 120-xxxxx
Summary of Contents for Stabil-Ion 370 Series
Page 2: ......
Page 8: ...Table of Contents iv Series 370 Stabil Ion Instruction Manual 370119...
Page 36: ...2 Initial Setup Procedures 2 16 Series 370 Stabil Ion Instruction Manual 370119...
Page 46: ...3 Installation 3 10 Series 370 Stabil Ion Instruction Manual 370119...
Page 58: ...3 Installation 3 22 Series 370 Stabil Ion Instruction Manual 370119...
Page 121: ......