3
Installation
3-13
Series 370 Stabil-Ion Instruction Manual - 370119
3.7.3 Using the Remote Outputs
The relay contact outputs are accessible on the 15 pin remote output connector. These contacts
provide status indication for IG1 and IG2 ON or OFF, degas ON or OFF, and fault detection. Figure
3-11 shows the pin designations. See Section 7.3 on page 7-4 for a description of fault relay
indications.
3.8 Connecting the RS-232 Computer Interface
Instructions for setting up this interface are in Section 2.1.4 on page 2-8.
This factory or field installed capability produces the signals shown in Table 3-4.
A mating DB-25S connector is supplied in the hardware kit. Use shielded cable to minimize
electromagnetic radiation or susceptibility. Ground the shield to the metal connector shell or to
Pin 1. Do not connect the shield to Pin 7.
Figure 3-11
Remote Output Connector (Use DA-15S Mating Connectors, Supplied)
Figure 3-12
Control Unit Rear Panel Showing RS-232 Connector
Remote Output
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
DA-15P
C
G
G
A
U
G
E
R
E
M
O
T
E
C
A
X
W
8
1
9
15
9
26
1
19
25
13
1
14
9
1
15
8
13
25
1
14
ANALOG
OUTPUT B
ANALOG
RS232 (or IEEE488)
ELECTROMETER
PROCESS CONTROL
REMOTE
OUTPUT
POWER
SUPPLY
GND
ANALOG
OUTPUT A
IG2 COL.
CONVECTRON
OUTPUT
3
7
0
1
0
4
IG1 COL.
DB-25P connector
Summary of Contents for Stabil-Ion 370 Series
Page 2: ......
Page 8: ...Table of Contents iv Series 370 Stabil Ion Instruction Manual 370119...
Page 36: ...2 Initial Setup Procedures 2 16 Series 370 Stabil Ion Instruction Manual 370119...
Page 46: ...3 Installation 3 10 Series 370 Stabil Ion Instruction Manual 370119...
Page 58: ...3 Installation 3 22 Series 370 Stabil Ion Instruction Manual 370119...
Page 121: ......