Product Location and Requirements
Chapter Two: Installation
22
Unpacking Checklist
Standard Equipment:
•
Type 870/872 Unit
•
Type 870/872 Instruction Manual (this book)
Optional Equipment:
•
Electrical Connector Accessories Kit (includes mating connectors for the connector
option, should you choose to make your own interface cable):
870A-K1
872A-K1
•
Interface cables (refer to Table 6, page 24)
•
MKS Local Display Module (Type LDM)
Product Location and Requirements
Operating Temperature
The acceptable operating temperature for the 870/872 unit is 0° to 50° C (32° to 122° F).
Power Requirements
The 870/872 transducer requires a minimum power supply of 12 VDC and a maximum power
supply of 36 VDC, regulated to ±5%. The specific input power required depends on the output
configuration of your unit, as listed in Table 5. The power is introduced to the 870/872 transducer
through the connector attached to the cable on top of the unit.
Power Requirements
Output
Configuration
Output
Signal
Input Power Requirement
Voltage
0 to 5 V
+12 VDC to +32 VDC (regulated if below 13 VDC)
@ 10 mA maximum
Voltage
0 to 10 V
+13 VDC to +32 VDC @ 10 mA maximum
Current
4 to 20 mA
+13 VDC to +36 VDC excitation (32 VDC maximum
across the transducer)
Table 5: Power Requirements
Summary of Contents for Micro-Baratron 870B
Page 3: ...123655 P1 Rev B 06 01 MKS Type 870B 872B Micro Baratron Absolute Pressure Transducer...
Page 8: ......
Page 10: ......
Page 28: ......
Page 52: ...Labels Chapter Three Overview 42 This page intentionally left blank...
Page 66: ...Model Code Appendix B Model Code Explanation 56 This page intentionally left blank...