Select the “Use the following fixed settings” option.
Enter the IP address and the Subnet issued to you by your IT Specialist.
The “Default Gateway” is used in situations where multiple networks exist, or
for connection to the HPQ3
from outside its local network. Your IT Specialist
will instruct you on what to enter here, otherwise leave it blank.
Once the relevant information has been entered, click OK to store and
commit the changes to the HPQ3
.
6. Baking
The analyser may be baked up to 250
o
C with the electronics package (HPQ3) removed. Care
should be taken to avoid exposing the ceramic feedthrough to sudden changes in temperature.
The analyser can be baked to 200
o
C with the electronics package attached if the package is not
exposed to temperatures above 40
o
C.
If the analyser flange temperature exceeds 200
o
C the HPQ3 must be removed.