Chapter 5
88
Series 358 Micro-Ion Controller Instruction Manual - 358013 - Rev. B
Command Syntax for
RS-232 Computer
Interface
DG
Definition:
Turn degas ON or OFF.
Modifiers:
ON or OFF
Response:
OK if command accepted, or INVALID if
rejected.
Example:
From computer:
DG ON CRLF
From controller:
OKCRLF
• Command is INVALID if the Ion gauge is OFF.
• A response to the DG ON command of OK indicates only that a signal
requesting degas has been sent to the electrometer. Degas may fail to
activate if the pressure is above 5 x 10-5 Torr. Use the DGS command
(see below) to verify that degas has been successfully initiated.
DGS
Definition:
Display degas status.
Modifiers:
None
Response:
ASCII 1 if degas is ON, 0 if degas is OFF.
Example:
From computer:
DGS CRLF
From controller:
1CRLF (Indicating degas is ON.)
DS
Definition:
Display pressure reading.
Modifiers:
IG or CG1 or CG2.
Response:
ASCII string representing the pressure for the
selected gauge.
Example:
From computer:
DS IG CRLF
From controller:
1.20E-07CRLF
• The DS IG command will return pressure from the top display if either
filament is ON, and 9.90E+09 if the gauge is OFF.
• The DS CG1 command will return pressure from the middle display.
• The DS CG2 command will return pressure from the bottom display.
Summary of Contents for Granville-Phillips 358 series
Page 2: ......
Page 8: ...Table of Contents 8 Series 358 Micro Ion Controller Instruction Manual 358013 Rev B...
Page 26: ...Chapter 2 26 Series 358 Micro Ion Controller Instruction Manual 358013 Rev B...
Page 48: ...Chapter 3 48 Series 358 Micro Ion Controller Instruction Manual 358013 Rev B...
Page 94: ...Chapter 5 94 Series 358 Micro Ion Controller Instruction Manual 358013 Rev B...
Page 104: ...104 Series 358 Micro Ion Controller Instruction Manual 358013 Rev B...
Page 116: ...Chapter 8 116 Series 358 Micro Ion Controller Instruction Manual 358013 Rev B...
Page 120: ...120 Series 358 Micro Ion Controller Instruction Manual 358013 Rev B...
Page 121: ......