Chapter Four: Operation
How To Prepare for Maximum Bakeout
35
Disassembly
To remove the sensor housing:
1. Turn off power to the instrument.
2. Use a flat head screwdriver to loosen the two cable strain relief screws.
3. Use a Phillips screwdriver to loosen and remove the two top housing screws that hold the
metal housing over the sensor.
4. Use a Phillips screwdriver to loosen and remove the three bottom housing screws that
attach the housing to the faceplate.
5. Remove the housing.
6. Use a ¼” nut driver to loosen and remove the upper two standoffs at the top of the cable
clamping assembly.
7. Disengage the cable clamping assembly from the middle standoffs.
8. Use needle-nose pliers to gently lift out the pin connectors.
Caution
Be sure to grip the pin connectors and not the coaxial cables.
9. Use a ¼” nut driver to loosen and remove the middle standoffs.
10. Remove the entire cable assembly.
The sensor is ready for bakeout up to 200° C.
Note
It is not necessary to remove the faceplate assembly during bakeout.
Summary of Contents for Baratron 121A
Page 3: ...113733 P1 Rev C 7 98 MKS Baratron Type 121A Absolute Pressure Transducer...
Page 7: ...Table of Contents v Appendix D High Pressure Fitting Selection 49 Description 49 Index 51...
Page 8: ...Table of Contents vi...
Page 10: ...List of Figures and Tables viii...
Page 30: ...Customer Support Chapter One General Information 20 This page intentionally left blank...
Page 42: ...Labels Chapter Three Overview 32 This page intentionally left blank...
Page 52: ...General Information Chapter Six Maintenance 42 This page intentionally left blank...
Page 60: ...Description Appendix D High Pressure Fitting Selection 50 This page intentionally left blank...
Page 62: ...Index 52...