Safety Procedures and Precautions
Pressure Transducer Safety Information
4
CHECK FOR LEAK-TIGHT FITTINGS
Carefully check all vacuum component connections to ensure leak-tight installation.
OPERATE AT SAFE INLET PRESSURES
Never operate at pressures higher than the rated maximum pressure (refer to the product
specifications for the maximum allowable pressure).
INSTALL A SUITABLE BURST DISC
When operating from a pressurized gas source, install a suitable burst disc in the vacuum system
to prevent system explosion should the system pressure rise.
KEEP THE UNIT FREE OF CONTAMINANTS
Do not allow contaminants to enter the unit before or during use. Contamination such as dust,
dirt, lint, glass chips, and metal chips may permanently damage the unit or contaminate the
process.
ALLOW PROPER WARM UP TIME FOR TEMPERATURE-CONTROLLED UNITS
Temperature-controlled units will only meet specifications when sufficient time is allowed for
the unit to meet, and stabilize at, the designed operating temperature. Do not zero or calibrate the
unit until the warm up is complete.
Summary of Contents for Baratron 121A
Page 3: ...113733 P1 Rev C 7 98 MKS Baratron Type 121A Absolute Pressure Transducer...
Page 7: ...Table of Contents v Appendix D High Pressure Fitting Selection 49 Description 49 Index 51...
Page 8: ...Table of Contents vi...
Page 10: ...List of Figures and Tables viii...
Page 30: ...Customer Support Chapter One General Information 20 This page intentionally left blank...
Page 42: ...Labels Chapter Three Overview 32 This page intentionally left blank...
Page 52: ...General Information Chapter Six Maintenance 42 This page intentionally left blank...
Page 60: ...Description Appendix D High Pressure Fitting Selection 50 This page intentionally left blank...
Page 62: ...Index 52...