Chapter Two: Installation
Setup
25
Setup
Follow the guidelines in this section when installing a 121 transducer.
Sensor
Connect the instrument to the vacuum system in either a horizontal or vertical (port down)
position, at the point where pressure is to be measured. Refer to Figure 1 for the sensor
dimensions.
Note
All dimensions are listed in inches with millimeters referenced in
parentheses.
Dimension A
2.560
(65.0)
2.50 (63.5)
Diameter
Swagelok
®
8-VCR
®
coupling
0.50 (12.7)
Figure 1: Sensor Dimensions
Dimension A on the Sensor
Full Scale Range
Dimension
1 Torr to 1K Torr
1.92 (48.4)
5K Torr to 25K Torr
2.02 (51.3)
Dimension listed in inches with millimeters referenced in parentheses
Table 6: Dimension A on the Sensor