Environmental Requirements
Chapter Two: Installation
24
Environmental Requirements
Follow these requirements when installing and using a Type 121 Absolute Pressure Transducer.
1. The normal operating temperature range, for the sensor and cable, is 0° C to 150° C
(32° F to 302° F).
The Type 121 sensor is bakeable to 200° C (392° F) with the sensor housing and cable
removed.
2. The normal operating temperature range, for the electronics unit (signal conditioner), is
0° C to 50° C (32° F to 122° F).
3. The power input required is ±15 VDC (±5%) @ 35 mA.
4. The overpressure limit is 35 psia or 120% of sensor Full Scale, whichever is greater.
5. It is important that the unit be kept free from vibration, because the diaphragm, when not
subject to gas damping at low pressure, may become susceptible to resonance. The
1 Torr F.S. units are the most sensitive; any vibration that does exist should be isolated
from the unit with vibration isolation mounting and a bellows coupling (MKS-supplied
or MKS-compatible).
For additional Type 121 transducer requirements refer to Appendix A: Product Specifications,
page 43.
Summary of Contents for Baratron 121A
Page 3: ...113733 P1 Rev C 7 98 MKS Baratron Type 121A Absolute Pressure Transducer...
Page 7: ...Table of Contents v Appendix D High Pressure Fitting Selection 49 Description 49 Index 51...
Page 8: ...Table of Contents vi...
Page 10: ...List of Figures and Tables viii...
Page 30: ...Customer Support Chapter One General Information 20 This page intentionally left blank...
Page 42: ...Labels Chapter Three Overview 32 This page intentionally left blank...
Page 52: ...General Information Chapter Six Maintenance 42 This page intentionally left blank...
Page 60: ...Description Appendix D High Pressure Fitting Selection 50 This page intentionally left blank...
Page 62: ...Index 52...