Chapter Four: Local Operation
Operation
65
5. Adjust the value of the analog set point to 100%.
The 4.5 Volt input to the Interface connector is now the full scale value. To be sure that
the analog set point is adjusted correctly, turn the adjust knob until the reading flickers
between 99.99 and 100.00%.
Note
1. The display reads 100.00% for an overrange reading as well as a true
100% reading.
2. If the current value of the analog set point exceeds
±
15% of full
scale, the controller will not change the full scale value.
6. Press the [STOP] key to return to the default
Pressure and Position
display screen.