Setup Messages
Chapter Five: Remote RS-232 Operation
96
How To Set the Process Limit Relays
The command [
P
x
value
] sets the process limit relay thresholds, where:
x:
1
= Low threshold - Process Limit 1
2
= High threshold - Process Limit 1
3
= Low threshold - Process Limit 2
4
= High threshold - Process Limit 2
value:
Pressure limit
To report a process limit threshold, issue the request:
R
xx
where
xx
:
11 = Low threshold - Process Limit 1
12 = High threshold - Process Limit 1
13 = Low threshold - Process Limit 2
14 = High threshold - Process Limit 2
The controller responds with the message [
P
x value
], where:
x
:
1
= Low threshold - Process Limit 1
2
= High threshold - Process Limit 1
3
= Low threshold - Process Limit 2
4
= High threshold - Process Limit 2
value
:
Pressure limit
To report the low threshold for process limit 1, enter:
R
10
An example response, if the low threshold for process limit 1 is 100 Torr, is:
P
1 100
Note
To disable a low threshold process limit, set the
P
x value
to negative full
scale.
To disable a high threshold process limit, set the
P
x value
to full scale.
Summary of Contents for 651D Series
Page 1: ...118965 P1 Preliminary 5 02 MKS Type 651D Pressure Controller...
Page 8: ......
Page 10: ......
Page 12: ......
Page 20: ...Customer Support Chapter One General Information 8 This page intentionally left blank...
Page 50: ...Labels Chapter Three Overview 38 This page intentionally left blank...
Page 78: ......
Page 149: ...Appendix C Initial Settings 137 This page intentionally left blank...
Page 150: ......
Page 158: ...Appendix E RS 232 Message Summary 146 This page intentionally left blank...