Chapter 3
46
Micro-Ion
®
Module Instruction Manual - 356007-GP
Table 3-14 RS-485 Command Set for Commands Affecting Module Operation
Command
Command
Type
Non-volatile
Default
Change
After
Reset
Data Returned
Can be
Locked
See
Page #
DG
Read/write
No
OFF (0)
No
Confirm or state
No
FAC
Write
Yes
Factory defaults
Yes
Confirm
No
IDT
Read/write
Yes
0 seconds
No
Confirm or state
No
IG
Read/write
No
ON (1)
No
Confirm or state
No
IGM
Read/write
Yes
ON (1)
No
Confirm or state
No
KBL
Write
Yes
Keyboard is unlocked
No
Confirm
No
KBU
Write
Yes
Keyboard is unlocked
Yes
Confirm
No
PC
Read/write
Yes
0.0
No
Confirm or state
No
PCH
Read/write
Yes
Relay deactivates at
10% above activation
pressure
No
Confirm or state
No
PCP
Read/write
Yes
Relay activates with
decreasing pressure
No
Confirm or state
No
RST
Write
No
None (not applicable)
No
None
No
SA
Write
Yes
00
Yes
Confirm
Yes
SB
Write
Yes
19200 baud
Yes
Confirm
Yes
SER
Read/write
Yes
1.0 x 10
–5
Torr
1.33 x 10
–5
mbar
1.33 x 10
–3
pascal
No
Confirm or state
Yes
SF
Read/write
Yes
• Manual (MAN) mode
for tungsten filaments
• Alternating (ALT)
mode for yttria-coated
iridium filaments
No
Confirm or state
Yes
SP
Write
Yes
8 data bits
No parity
Yes
Confirm
Yes
SU
Write
Yes
Torr, mbar, or pascal as
specified by customer
Yes
Confirm
Yes
TLU
Write
Yes
Functions unlocked
No
Confirm
No
TS
Write
Yes
Factory setting
No
Confirm
No
TZ
Write
Yes
Automatically set
No
Confirm
No
UNL
Write
Yes
Functions unlocked
No
Confirm
No